Owner's Manual

11
Joerns® Bed Frames
WeCare™ Bed
© 2021 Joerns Healthcare • 6110429 RevC • 21-2257
To Remove Actuator
Recommended Tools: Screwdriver, Hammer, Punch
1. 
2. Tip bed on its side or upside down to replace hi-lo
actuators.
3. Unplug actuator cable from actuator.
4. 
screwdriver to remove clip on each pin.
5. 
and punch to tap out the pins.
6. 
 
making note of the following two important items:

Back and hi-lo actuators also include washers.
Note: Ensure all hardware is re-installed in the same
locations.
To Remove AC Power Cord
Recommended Tools: Small Screwdriver, Masking
Tape, Cable Ties
1. Raise bed to maximum height if possible.
2. Unplug power cord from wall.
3. 
Tape remaining cables in their original position. It




Warning: Possible Shock Hazard.


to check condition of power cord on a regular basis.

4. 
5. 



connects to control box until plug comes out.
6. Plug replacement cord into control box until it
latches.
 
position as original routing.
 Replace cable ties along with cables in their
original position and routing to the control box.


Troubleshooting

with the following steps:
Event Possible Remedies
Bed not
responding
to pendant
control
1. 
inserted with cables secured.


2. 
connected and secure.
3. 
power source and control box
green light is illuminated
4. Thermal overload condition. This
should self-correct in 2 minutes

󰀨
5. 
and clear of obstructions
6. If these steps fail to resolve the


Back sleep
surface or
knee surface
does not
raise or lower
1. Ensure the actuator pins and


2. 

the correct location.
Bed moves
across the

height
1. 
2. 

Note: Electrical damage can occur if a non-approved

To Maximize Service Life
1. Keep the bed clean.
2. 


3. 





appropriate service parts.
4. Make sure each inspection includes the underside
of the bed frame and mattress support platform.

loose fasteners.