Position the resident’s head in the position that
the resident will be in for the largest portion of the
day. If the resident is lying at, please use three
(3) ngers for the hand check. If the resident will
be sitting up for the majority of the day, please use
two (2) ngers.
Perform a hand check by placing ngers locally
under the resident’s buttocks between two
cushions. The resident should not bottom out. If
they do, increase the therapy control by one LED
(light), until they no longer
bottom out.
The safety side bolsters cannot be inated unless
the CPR plug is installed. The plug is located
on the hose assembly between the mattress
and the control unit. Caution: The safety side
bolsters should always be inated when resident is
unattended on the mattress and/or the mattress is
in Turn mode.
Operation
Warning: For important precautions, please see
page two.
Caution: The resident’s head should be positioned
in the center of the top section of the mattress. When
using the mattress system always ensure that the
resident is positioned properly within the connes of
the bed. Do not let any extremities protrude over the
side or between the bed rails when the mattress is
being used.
Resident Comfort Controls and Monitoring
Power
The Power button is used to turn the power on and off.
Autorm
Autorm mode provides maximum air ination
designed to assist both residents and caregivers
during resident transfer and treatment. The unit will
automatically return to the mode it was in prior to
Autorm (either Turn or Static) in approximately
15 minutes.
Pause
Use this feature to immediately stop all rotation and
hold the current mattress position. The mattress will
hold this position until this feature is turned off by
pressing the Pause button a second time.
8
Lockout
This unit is designed to lock out all of the adjustment
controls after the resident has been positioned
correctly. Ten minutes after the last button has been
pushed, push the Power light begins to ash indicating
Lockout is enabled. This feature is to prevent any
unauthorized changes to the resident settings. To
unlock and make adjustments to the settings, press
both up and down comfort arrows at the same time.
Lockout mode will return after ten minutes.
Note: The unit is designed to lock out all the
adjustment controls after the resident has been
positioned correctly. In approximately ten minutes after
the last button push the power on light begins to ash
indicating Lockout is enabled.
Comfort Adjust
The Comfort Adjust function is located across the
top of the control panel. The DermaFloat
®
LR can be
customized to meet individual resident needs within
a therapeutic window. Use the soft and rm keys to
simultaneously increase or decrease pressure in the
entire mattress in Turn and Static modes. The current
setting is shown in the LED display in the center of the
control panel.
Turn Mode
The Turn mode and sets the desired rotation type. The
resident can be rotated to either side individually or to
both sides in a continuous cycle.
Turn Time
Use the Turn Time feature to set the time cycle for
the selected Turn mode. The DermaFloat LR can be
placed into a 10, 20, or 60 minute cycle based on what
is determined to be the optimal therapy setting for the
individual resident.
Figure 3