User-Service Manual Joerns® Bed Frames WeCare™ Bed To avoid injury, read user’s manual before using.
Joerns® Bed Frames WeCare™ Bed Important Precautions power cord. Always follow the proper factory-installed routing configuration. Failure to do so could result in injury or death. Warning: Possible Injury Or Death. This product is intended for use as an adjustable mattress support platform that contours for resident/patient comfort with adjustable height for resident/patient and caregiver convenience.
Joerns® Bed Frames WeCare™ Bed Warning: Possible Injury Or Death. Do not use an assist device if any openings within assist device body allow resident/patient to get his/her head or neck lodged within these openings. Use of assist devices with large openings, may result in injury or death. Warning: Possible Injury Or Death. Do not use any Joerns® assist device until you verify it is designed specifically for the WeCare bed and it is locked in place.
Joerns® Bed Frames WeCare™ Bed Table of Contents Important Precautions .................................................................................................................................................. 2 Entrapment Information ................................................................................................................................................ 4 Features and Specifications .............................................................................................
Joerns® Bed Frames WeCare™ Bed Features and Specifications Joerns® WeCare™ Bed Standard Features 80" Mattress Support Platform 8" to 28" Travel Range Quiet DC Actuators Welded Tubular Frame Grid Sleep Surface Wide Stable Base SoftTone Frame Color Heavy-Duty Casters Locking System: • Roll-in-Low Lock System • Roll at any Height Mobility Baseboard Bumper Mattress Stops Electrical Specifications Power: 120V Frequency: 60Hz Current Rating: 4.
Joerns® Bed Frames WeCare™ Bed Assembly Caution: Carefully remove the packaging straps (orange zip ties) before attempting to operate the bed. Caution: Prior to operation, refer to operating conditions above to ensure bed electronics are within ambient temperature. Check that all parts of the bed were delivered and inspect for damage. Parts to Prepare: a. Head End Frame b. Foot End Frame c. Safety Pin (2) d. Actuator Pin and Clip (2 or 4) e. Pendant f.
Joerns® Bed Frames WeCare™ Bed 6. Line up back and knee actuators and insert actuator pin then add clip to lock in place (Figure 7 and 8). Ensure the washers are used on the back actuator (Figure 9). 7. Locate back actuator and head end hi-lo cables and route through safety pin bail (Figure 6), then route behind cable bundle (Figure 10 and 11). 8. Insert back actuator cable in port 1 and insert hi-lo actuator head end cable in port 3 (Figure 12). 9. Secure cables with cable clamp (Figure 13). 10.
Joerns® Bed Frames WeCare™ Bed Baseboard Bumper Recommended Tool: Rubber mallet 1. Place the baseboard bumper as shown (Figure 14). 2. Tap into place with a rubber mallet or press firmly with the sole of your shoe until it locks into place as shown (Figure 15). Figure 14 Note: If a longer baseboard bumper is needed for use with Trendelenburg/Reverse Trendelenburg, recommend ordering bumper 11011899AL. Mattress Stops Warning: Possible Injury.
Joerns® Bed Frames WeCare™ Bed Installation Of Head and Foot Panels Refer to Bed Panel Installation Sheet for more Tools required: #2Phillips screwdriver Head Panel Installation: 1. Head panel has label on back. 2. Lay panel on soft surface front face down. 3. Head panel brackets are longer of two sizes (Figure 19). 4. Install head panel brackets with four screws provided. Bracket offset should face up (Figure 20). Do not tighten until after fitting mounting brackets into bed sockets. 5.
Joerns® Bed Frames WeCare™ Bed For models with staff control in foot panel - cable install and routing: 1. Raise the bed to a comfortable working height and raise the foot section with the pendant. Then unplug the bed. 2. Locate the two staff control cables. Lay them out side by side with one cable reversed. A male and female connection should be at each end. 3. Unplug the pendant. Route one cable under the control box and plug into pendant port.
Joerns® Bed Frames WeCare™ Bed Operation Pendant Operation Warning: Possible Injury or Death. This bed should not be operated (control the movements of the bed through the control interfaces) by persons who do not have the cognitive skills to understand the information in the owner’s manual or cannot understand and perform the proper operation of the bed. This bed should not be used or operated by children. Back Operation 1. Push button to raise back 2.
Joerns® Bed Frames WeCare™ Bed Caution: Be careful not to damage power cord when moving or placing bed. After placing bed in service, monitor power cord positioning carefully to prevent damage as you fully operate all bed functions. Caution: When power cord is not plugged into a wall outlet, place it on sleep surface to prevent it from contacting floor. This will help prevent cord damage when transporting bed. Caution: If a power cord is damaged, remove bed from service until repairs are performed.
Joerns® Bed Frames WeCare™ Bed Actuator and Control Box Troubleshooting Any loss of electrical function should be investigated with the following steps: 1. Make sure that all cables are plugged in, including pendant cable and AC power cord. 2. Switch pendant with a new one or one from a bed that is functioning normally. 3. If these steps fail to resolve the problem, call Technical Support at 800-826-0270. 6. Plug replacement cord into control box until it latches. 7.
Joerns® Bed Frames WeCare™ Bed To Maximize Service Life Do not submerge the pendant in liquid. 1. Keep the bed clean. 2. Observe bed operation and report any discrepancies or problems to Joerns Healthcare Technical Support at 800.826.0270 3. After initial week of use, check all threaded fasteners for looseness, and make sure all pins are in their normal location and fastened securely. Check monthly for loose bolts, nuts, pins and other retaining hardware.
Joerns® Bed Frames WeCare™ Bed Accessories Trapeze And Trapeze Adapter Accessory Weights Warning: Possible Injury Or Death. Improper installation of the trapeze adapter could alter or dislodge the head panel position, increasing the risk of injury or suffocation. Drainage Bag Holder: 0.85 lbs (0.38 kg) F028 Side Assist (pair): 16.5 lbs (7.48 kg) F14SC Side Assist (pair): 13.7 lbs (6.21 kg) * Head / Foot Board ~ 14.9 lbs (6.76 kg) * Mattress ~ 24.0 lbs (10.
Joerns® Bed Frames WeCare™ Bed Trapeze Adapter (Figure 34) 1. The trapeze adapter is designed to allow attachment of the trapeze system to the WeCare™ Bed. Prior to use, please read important warnings related to this option and follow all instructions below. 2. Install the trapeze adapter according to the supplied instructions. If needed, call Technical Support at 800-826-0270 for instructions. 3. Install the trapeze assembly onto the trapeze adapter according to the trapeze manufacturer’s instructions. 4.
Joerns® Bed Frames WeCare™ Bed Appendix A: Electromagnetic Compatibility (EMC) Information Manufacturer’s Declaration –Electromagnetic Emissions The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such an environment.
Joerns® Bed Frames WeCare™ Bed Guidance and Manufacturer’s Declaration - Electromagnetic Immunity: The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such an environment.
Joerns® Bed Frames WeCare™ Bed Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Joerns WECC bed is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or the user of the Joerns WECC bed should assure that it is used in such and environment.
Joerns Healthcare Warranty Program for Joerns® WeCare™ Bed Frame Joerns® WeCare™ Bed is guaranteed for a period of one year from the date of delivery, against defects in materials and workmanship, under normal use and service. Joerns Healthcare’s obligation under this warranty is limited to supplying replacement parts, servicing or replacing, at its option, any product that is found by Joerns Healthcare to be defective. This one-year warranty includes all mechanical and electrical components.