User`s guide

8
To Test Cable from One End
This can be done immediately after termination
of one end of the cable, because no connector
or remote is required at the other end of the
cable. Attach a remote after the pre-test and
you are ready for a full test when the other end
is terminated.
1) Attach one end of supplied one-foot jumper
to main unit and the other end to the wall
plate or patch panel jack.
2) Turn on unit by pressing any button. If not
in pre-test mode, press MODE until TEST/
PRE-TEST is displayed, than press
SEL button.
3) Results should appear within 5 seconds.
Pressing SEL button starts a new test
immediately.
4) Disconnect cable after test. The test
repeats every 5 seconds.
Application Hints: The jumper cables must be
short compared to the cable run for accurate
open and split pair indication, either no more
than 10% of the total run length or 3 feet,
whichever is less. Cable being tested must be
at least 4 feet long for pre-test to work properly.