www.janitza.com Art. Nr. 33.03.162 Dok. Nr. 1.040.071.1.m Power Analyser Janitza electronics GmbH Vor dem Polstück 1 D-35633 Lahnau Support tel. 0049 6441 9642-22 Fax 0049 6441 9642-30 E-mail: info@janitza.com Internet: http://www.janitza.
UMG 96RM-M Contents General Incoming goods inspection Scope of delivery of the UMG 96RM-M Available accessories Product description Intended use Characteristics of the UMG 96RM-M Measuring method Netzanalysesoftware GridVis Connection options Assembly Installation Supply voltage Voltage metering Current measurement M-Bus interface Digital outputs Operation Display mode Programming mode Parameters and measured values Configuration Applying the supply voltage Current and voltage transformers Programming curre
UMG 96RM-M Digital outputs 66 Pulse output 68 Comparator 74 Parameter list comparator and digital outputs 77 Service and maintenance 80 Device calibration 80 Calibration intervals 80 Error messages 82 Technical data 88 Parameters of functions 94 Table 1 - Parameter list 96 Dimensional drawings 102 Overview of measured value displays 104 Declaration of conformity 110 Anschlussbeispiel 111 Brief instructions 112 3
UMG 96RM-M General Comments about the manual Copyright Your comments are welcome. If anything in this manual is unclear, please let us know and send us an e-mail at: info@janitza.com This manual is subject to the laws of copyright protection and may not be mechanically or electronically photocopied, reprinted, reproduced or otherwise reproduced or published in part or as a whole, without the legally binding, written consent of Janitza electronics GmbH, Vor dem Polstück 1, D 35633 Lahnau, Germany.
UMG 96RM-M Application notes Please read these operating instructions and all other publications that must be consulted in order to work with this product (particularly for installation, operation or maintenance). When using the device, the legal and safety regulations required for the respective application must also be observed. Please observe all safety regulations and warnings. Noncompliance with the instructions can lead to personal injury and/or damage to the product.
UMG 96RM-M About these operating instructions Incoming goods inspection These operating instructions are part of the product. • Read the operating instructions prior to using the device. • Keep the operating instructions at hand throughout the entire service life of the product and keep ready for referencing. • Hand over the operating instructions to each subsequent owner or user of the product.
UMG 96RM-M Scope of delivery of the UMG 96RM-M Quantity Item no. Designation 1 52.22.039 UMG 96RM-M 2 29.01.036 Mounting brackets 1 33.03.162 Operating instructions 1 51.00.116 CD with the following contents - GridVis programming software - GridVis functional description 1 10.01.855 Screw terminal, pluggable, 2-pin (auxiliary energy) 1 10.01.849 Screw terminal, pluggable, 4-pin (voltage measurement) 1 10.01.871 Screw terminal, pluggable, 6-pin (current measurement) 1 10.01.
UMG 96RM-M Product description Intended use The UMG 96RM-M is provided for the measurement and calculation of electrical parameters such as voltage, current, power, energy, harmonics, etc. for building installations, to distributors, circuit breakers and busbar trunking systems. The UMG 96RM-M is suitable for installation in permanent, weatherproof switchboards. Conducting switchboards must be earthed. Measurement voltages and measurement currents must originate from the same grid.
UMG 96RM-M Characteristics of the UMG 96RM-M • General • Front panel-mounted with the dimensions 96x96 mm • Connection via screw-type terminals • LC display with backlighting • Operation via 2 buttons • 3 voltage measurements inputs (300V CATIII) • 3 current measurement inputs for current transformer • M-Bus interface • 2 digital outputs • Working temperature range -10°C ..
UMG 96RM-M Measuring method The UMG 96RM-M measures uninterrupted and calculates all root mean squares over a 10/12-period interval. The UMG 96RM-M measures the true root mean square (TRMS) of the voltages and currents applied to the measuring inputs. Operating concept The UMG 96RM-M can be programmed directly on the device via the 2 buttons. In addition, measurement values can be called up via the M-Bus interface - e.g. with the GridVis read-out software.
UMG 96RM-M Netzanalysesoftware GridVis Connection options The UMG 96RM-M can be programmed and read with the GridVis network analysis software which is part of the scope of delivery. For this a PC must be connected via a serial interface (RS232 / USB) for example via an M‑Bus Master (level converter) to the M-Bus interface of the UMG 96RM-M. Connection of a UMG 96RM-M to a PC via a M-Bus signal converter (RS232): PC RS232 GridVis Signal converter M-Bus UMG 96RM-M Item-No. 15.06.
UMG 96RM-M Assembly Installation location Mounting The UMG 96RM-M is suitable for installation in permanent, weatherproof switchboards. Conducting switchboards must be earthed. The UMG 96RM-M is mounted on the switchboard by the side mounting brackets. These must be removed before using the device. Mounting is carried out by inserting and engaging the brackets. Installation position The UMG 96RM-M must be installed vertically in order to achieve sufficient ventilation.
UMG 96RM-M 13
UMG 96RM-M Installation Supply voltage L A supply voltage is required to operate the UMG 96RM‑M. N The voltage supply is connected via plug-in terminals on the back of the device. Before applying the supply voltage, ensure that the voltage and frequency correspond with the details on the nameplate! Fuse Separator The supply voltage must be connected via a UL/IEC approved fuse (6 A, type C). Fig.
UMG 96RM-M m • In building installations, the supply voltage must be provided with a disconnect switch or circuit breaker. • The disconnect switch must be attached near the device and must be easily accessible by the user. • The switch must be labelled as a separator for this device. • Voltages that exceed the permissible voltage range can destroy the device.
UMG 96RM-M Voltage metering The UMG 96RM-M can be used for voltage measurement in TN, TT and IT systems. Voltage measurement in the UMG 96RM-M is designed for the 300 V overvoltage category CATIII (4 kV rated pulse voltage). In systems without a neutral, measured values that require a neutral refer to a calculated neutral. L1 L1 277V/480V 50/60Hz L2 480V 50/60Hz L2 L3 L3 N PE Impedanz V1 V2 V3 VN V2 V1 V3 VN Auxiliary energy Fig.
UMG 96RM-M Rated mains voltage Lists of the networks and their rated mains voltage in which the UMG 96RM-M can be used. Three-phase 4-wire systems with earthed neutral conductor. Unearthed three-phase, 3-wire systems. UL-N / UL-L UL-L 66 V/115 V 120 V/208 V 127 V/220 V 220 V/380 V 230 V/400 V 240 V/415 V 260 V/440 V 277 V/480 V 66 V 120 V 127 V 220 V 230 V 240 V 260 V 277 V 347 V 380 V 400 V 415 V 440 V 480 V Maximum rated voltage of the network Fig.
UMG 96RM-M Voltage measurement inputs The UMG 96RM-M has three voltage measurement inputs (V1, V2, V3). L1 L2 Overvoltage The voltage measurement inputs are suitable for measurement in networks in which overvoltages of overvoltage category 300V CATIII (4 kV rated pulse voltage) can occur. Frequency The UMG 96RM-M requires the mains frequency for the measurement and calculation of measured values. The UMG 96RM-M is suitable for measurements in the frequency range of 45 to 65 Hz. L3 N Fuse Separator Fig.
UMG 96RM-M When connecting the voltage measurement, the following must be observed: • A suitable separator must be provided in order to switch off the power to the UMG 96RM-M. • The separator must be placed near the UMG 96RMM, marked for the user and easily accessible. • Use a fuse protected, UL/IEC approved 10A circuit breaker (type C) as an overcurrent protection device and separator.
UMG 96RM-M Connection diagram, voltage measurement • 3p 4w (addr. 509= 0), factory setting • 3p 4wu (addr. 509 = 1) L1 L1 L2 L2 L3 N L3 N V1 V2 V3 VN Fig. System with three-phase conductors and a neutral conductor. • 3p 4u (addr. 509 = 2) V1 L1 L2 L3 L3 V3 VN Fig. System with three-phase conductors and no neutral conductor. Measured values that require a neutral refer to a calculated neutral. 20 VN • 3p 2u (addr. 509 = 5) L2 V2 V3 Fig.
UMG 96RM-M • 1p 2w1 (addr. 509 = 4) • 2p 4w (addr. 509 = 3) L1 L1 L2 L3 N N V1 V2 V3 V1 VN Fig. Measured values derived from the V2 and V3 voltage measurement inputs are assumed to be zero and not calculated. • 1p 2w (addr. 509 = 6) V2 V3 VN Fig. System with uniform phase loading. The measured values for the V2 voltage measurement input are calculated. • 3p 1w (addr. 509 = 7) L1 L1 L2 L3 L2 L1 L2 L3 V1 V2 V3 VN Fig. TN-C system with single-phase, three-wire connection.
UMG 96RM-M Current measurement The UMG 96RM-M is designed for connecting current transformers with secondary currents of ../1A and ../5A. The factory set current transformer ratio is 5/5 A and may need to be adapted to the current transformers. It is not possible to perform a direct measurement without a current transformer with the UMG 96RM-M. Only AC currents (and not DC currents) can be measured. c m c 22 Attention! The current measurement inputs are dangerous to touch.
UMG 96RM-M Direction of the current If incorrectly connected, a subsequent re-connection of the current transformer is required.
UMG 96RM-M Connection diagram, current measurement • 3p 4w (addr. 510= 0), factory setting • 3p 2i (addr. 510 = 1) L1 L1 L2 L2 L3 N L3 N I1 I2 I3 Fig. Measurement in a three-phase net-work with an unbalanced load. • 3p 2i0 (addr. 510 = 2) I1 • 3p 3w3 (addr. 510 = 3) L1 L2 L2 L3 L3 I2 I3 Fig. The measured values for the I2 current measurementinput are calculated. 24 I3 Fig. System with uniform phase loading. The measured values for the I2 current measurement input are measured.
UMG 96RM-M • 3p 3w (addr. 510 = 4) • 2p 4w (addr. 510 = 5) L1 L1 L2 L2 L3 N L3 N I1 I2 I3 Fig. System with uniform phase loading. The measured values for the I2 and I3 current measurement inputs are calculated. • 1p 2i (addr. 510 = 6) L1 I1 I2 I3 Fig. System with uniform phase loading. The measured values for the I2 current measurement input are calculated. • 1p 2w (addr. 510 = 7) L1 L2 N I1 I2 I3 Fig.
UMG 96RM-M Connection diagram, current measurement • 3p 1w (addr. 510 = 8) L1 L2 L3 L1 L2 L3 L1 L2 L3 I1 I2 I3 Fig. Three systems with uniform phase loading. The current measurement values of the phases of the respective system where are no CTs connected are calculated (I2/I3 resp. I1/I3 resp. I1/I2). 26 c Caution! The UMG96RM-M is only approved for a current measurement using the current transformer.
UMG 96RM-M Total current measurement If the current measurement takes place via two current transformers, the total transformer ratio of the current transformer must be programmed in the UMG 96RM-M. UMG The UMG 96RM must then be set as follows: Primary current: 1000 A + 1000 A = 2000 A Secondary current: 5A I S1 S2 Einspeisung 1 Supply 1 1P1 (K) (L) 1P2 1S1 (k) (l) 1S2 Verbraucher A Consumer A P1 1S1 1S2 Example: The current measurement takes place via two current transformers.
UMG 96RM-M Ammeter If you want to measure the current not only with the UMG 96RM-M but also with the ammeter, the ammeter must be connected in series with the UMG 96RM-M. UMG S1 I S2 A Einspeisung Supply (k)S1 S2(l) (K)P1 P2(L) Verbraucher Consumer Fig. Current measurement with an additional ammeter (example).
UMG 96RM-M M-Bus interface Cable connections The M-Bus interface is designed with the UMG 96RM-M as a 2-pole plug contact and communicates via the MBus protocol. Twisted screened cable should be used for connections via the M-Bus interface. The UMG 96RM-M loads the M-Bus with an M-Bus device load of 1.5 mA. • Cable paths should be designed to be as short as possible. • Maintain as much distance as possible to power cables and to consumers (e.g. electrical motors, neon tubes, transformers).
UMG 96RM-M Cable type Bus structure The cable used must be suitable for an ambient temperature of at least 80 °C. • All devices are connected in a star, line or tree structure, whereby each device has its own address within the bus (see also Parameter programming). • A subdivision of the network structure into individual segments is implemented via repeaters (line amplifiers). • Up to 250 subscribers can be connected together in a single segment.
UMG 96RM-M Tree structure • This topology combines the star and line structures. Repeaters generally divide the branches into individual segments. Thus in the event of a fault only a specific branch is affected and so a fault in the bus system can be quickly localised.
UMG 96RM-M Digital outputs Digital outputs 1-2 Digital outputs 1-2 The digital outputs can switch DC and AC loads. The digital outputs are not short circuit protected. Connected cables longer than 30 m must be shielded. An external auxiliary voltage is required. The digital outputs can be used as pulse outputs. The digital outputs can output results from comparators. 13 Digital output 1 AC/DC 14 Digital output 2 K2 • • • • • • ~ ~ UMG 96RM-M AC/DC 15 K1 The UMG 96RM-M has 2 digital outputs.
UMG 96RM-M 33
UMG 96RM-M Operation The UMG 96RM-M is operated using buttons 1 and 2. Measured values and programming data appears on a liquid crystal display. A distinction is made between display mode and programming mode. The accidental changing of programming data is prevented by the entry of a password. Display mode In the display mode, you can scroll between the programmed measured value displays using buttons 1 and 2. All factory-set measured value displays listed in section 1 can be called up.
UMG 96RM-M Mean value Programming mode Sum measurement Phase conductorPhase conductor Password CT: Current transformer VT: Voltage transformer K1: Output 1 K2: Output 2 Button 2 Export Button 1 35
UMG 96RM-M Parameters and measured values Example of the parameter display All parameters necessary for operating the UMG 96RMM, e.g. the current transformer data, and a selection of frequently required measured values are stored in the table. The contents of most addresses can be accessed via the serial interface and the buttons on the UMG 96RM-M. On the UMG 96RM-M display the value “001” is shown as the content of address “000”.
UMG 96RM-M Button functions Display mode Programming mode Change mode Change mode Password simultaneous simultaneous Browse Browse short long short Measured values 1a Measured values 2a long Programming menu 1 Programming menu 2 long Measured values 2b short Programming menu 3 Programming Programming menu 1 Confirm selection Short: digit +1 Long: digit -1 (flashes) (flashes) Short: value x 10 (decimal to the right) Long: Value /10 (decimal to the left) 37
UMG 96RM-M Configuration Applying the supply voltage To configure the UMG 96RM-M, the supply voltage must be connected. The level of supply voltage for the UMG 96RM-M can be found on the nameplate. If no display appears, check the operating voltage to determine whether it is within the rated voltage range. Current and voltage transformers A current transformer is set to 5/5 A in the factory. The pre-programmed voltage transformer ratio only needs to be changed if voltage transformers are connected.
UMG 96RM-M Programming current transformers Switching to programming mode • Simultaneously press buttons 1 and 2 in order to switch to programming mode. If a user password was programmed, the password request will appear with "000". The first digit of the user password flashes and can be changed with button 2. The next digit is selected by pressing button 2 and will begin flashing. If the correct combination was entered or if no user password was programmed, the device will enter programming mode.
UMG 96RM-M Programming voltage transformers • Switch to the programming mode as described. The symbols for the programming mode (PRG) and for the current transformer (CT) appear. • Use button 2 to switch to the voltage transformer setting. • Confirm the selection with button 1. • The first digit of the input area for the primary current starts flashing.
UMG 96RM-M Programming parameters Switching to programming mode • Switch to the programming mode as described. The symbols for the programming mode (PRG) and for the current transformer (CT) appear. • Use button 2 to switch to the voltage transformer setting. The first parameter of the parameter list is shown by repeatedly pressing button 2. Changing parameters • Confirm the selection with button 1. • The most recently selected address is displayed with the associated value.
UMG 96RM-M Device address (addr. 000) If several devices are connected to one another via the M-Bus interface, a master device can only differentiate between these devices by means of their device addresses. Therefore, each device in a network must have a different device address. Addresses can be set in the range from 1 to 250. C The adjustable range of the device address is between 0 and 255. The values 0 and 251 to 255 are reserved and may not be used. Secondary device address (addr.
UMG 96RM-M Baud rate (addr. 001) Mean value A common baud rate is adjustable for the M-Bus interfaces. The baud rate must be chosen to be a uniform value in the network. The parameter data bits (8), parity (even) and stop bits (1) are permanently set. Mean values are formed over an adjustable for the current, voltage and power measured The mean values are identified with a bar the measured value. The averaging time can be selected from a list fixed averaging times.
UMG 96RM-M Averaging method Minimum and maximum values After the set averaging time, the exponential averaging method used achieves at least 95% of the measured value. All measured values are measured and calculated every 10/12 periods. Minimum and maximum values are determined for most of the measured values. The minimum value is the smallest measured value that has been determined since the last reset. The maximum value is the largest measured value that has been determined since the last clearance.
UMG 96RM-M Mains frequency (addr. 034) For automatic ascertainment of the mains frequency, an L1-N voltage larger than 10Veff must be applied to the voltage measurement input V1. The mains frequency is then used to calculate the sampling rate for the current and voltage inputs. Adjustment range: 0, 45 .. 65 0= automatic frequency determination. The mains frequency is determined from the measurement voltage. 45..65 = fixed frequency The mains frequency is preselected.
UMG 96RM-M Energy meter Reading the active energy The UMG 96RM-M has energy meters for active energy, reactive energy and apparent energy.
UMG 96RM-M Harmonics Total Harmonic Distortion (THD) Harmonics are the integer multiple of a mains frequency. The voltage mains frequency for the UMG 96RM-M must be in the range between 45 and 65 Hz. The calculated voltage and current harmonics refer to this mains frequency. Harmonics up to 40x the mains frequency are recorded. THD is the ratio of the root mean square value of harmonics to the root mean square value of the mains frequency.
UMG 96RM-M Measured value relay Changeover time (addr. 039) All measured values are calculated every 10/12 periods and can be recalled once per second on the measured value displays. Two methods are available for retrieving the measured value displays: Adjustment range: 0 .. 60 seconds If 0 seconds are set, no changeover takes place between the measured value displays selected for the measured value relay. The changeover time applies for all display change profiles.
UMG 96RM-M Measured value displays User password (addr. 050) After return of the power supply, the UMG 96RM-M shows the first measured value panel from the current display profile. In order to keep the selection of measured values to be displayed arranged in a clear manner, only one part of the available measured values is preprogrammed for recall in the measured value display by default.
UMG 96RM-M Clear energy meter (addr. 507) Rotation field direction The active, apparent and reactive energy meters can only be cleared together. The rotation field direction of the voltages and the frequency of phase L1 are shown on the display. The rotation field direction indicates the phase sequence in three-phase systems. Usually there is a "clockwise spinning rotation field". The phase sequence at the voltage measurement inputs is checked and displayed in the UMG 96RM-M.
UMG 96RM-M LCD contrast (addr. 035) The preferred direction of viewing for the LCD is from "below". The user can adjust the LCD contrast of the LCD screen. It is possible to set the contrast in the range from 0 to 9 in steps of 1. 0 = characters are very light 9 = characters are very dark Factory default setting: 5 Backlight Operation backlight (addr. 036) The operation backlight is activated by pushing the appropriate button, or with a restart. Standby backlight (addr.
UMG 96RM-M Time recording Operating hours meter The UMG 96RM-M records the operating hours and the total running time of each comparator The operating hours meter measures the time for which the UMG 96RM-M records and displays measured values. The time of operating hours is measured with a resolution of 0.1 h and is displayed in hours. The operating hours meter cannot be reset. • where the time of operating hours is measured with a resolution of 0.
UMG 96RM-M Serial number (addr. 754) The serial number shown by UMG 96RM-M has 6 digits and is part of the serial number displayed on the nameplate. The serial number cannot be changed. Serial number display Serial number information on the nameplate: XX00-0000 Software release (addr. 750) The software for UMG 96RM is continuously improved and expanded. The software version in the device is marked with a 3-digit number, the software release. The user cannot change the software release.
UMG 96RM-M Commissioning Applying the measured current Applying the supply voltage The UMG 96RM-M is designed for connecting ../1 A and ../5 A current transformers. Only AC currents and not DC currents can be measured via the current measurement inputs. Short circuit all current transformer outputs except for one. Compare the currents displayed on the UMG 96RM-M with the applied current.
UMG 96RM-M Rotation field direction Checking the measurement Check the direction of the voltage rotation field on the measured value display of the UMG 96RM-M. Usually there is a "clockwise" spinning rotation field. If all voltage and current measurement inputs are correctly connected, the individual and sum power ratings are accurately calculated and displayed.
UMG 96RM-M Check the sum power ratings If all voltages, currents and power ratings for the respective phase conductor are correctly displayed, the sum power ratings measured by the UMG 96RM-M must also be correct. For confirmation, the sum power ratings measured by the UMG 96RM-M should be compared with the energy of the active and reactive power meters at the power feed.
UMG 96RM-M M-Bus interface Number of data points The data of the parameter list and measurement values list can be accessed via the M-Bus interface with the help of the primary or secondary address. Changing these values is not possible via the M-Bus. The number of data points to be transmitted for the RSP_UD2 telegram is defined via this address. The primary device address is factory preset to „1“.
UMG 96RM-M Measurement signal level Structure of the RSP_UD2 telegram The data transfer in the M-Bus network is implemented through modulation of the supply voltage, whereby the voltage for a high signal is 36 V and 24 V for a low signal. The slave device answers the master via the modulation of its current draw, whereby the high signal is 1.5 mA and the low signal is 11-20 mA. Byte 1 2 Name Start Length Length Start 3 4 Cont. 68 Byte 7 8 Name CI ID1 Cont.
UMG 96RM-M List of data points Data point Discription Unit Resolut. Device Format Byte 1 Real energy, without backstop dev. Wh 10 0 6 2 Real energy, obtained Wh 10 0 6 3 Real energy, supplied Wh 10 0 6 4 Reactive energy, inductive varh 10 1 6 5 Reactive energy, capacitive varh 10 1 6 6 Reac. energy, without backst. dev.
UMG 96RM-M List of data points Data point Discription Unit Resolut.
UMG 96RM-M Telegramm Data point Discription DIF DIFE DIFE DIFE VIF VIFE 1 Real energy, without backstop dev. 0x06 X X X 0x04 X 2 Real energy, obtained 0x86 0x10 X X 0x04 X 3 Real energy, supplied 0x86 0x20 X X 0x04 X 4 Reactive energy, inductive 0x86 0x40 X X 0x04 X 5 Reactive energy, capacitive 0x86 0x50 X X 0x04 X 6 Reac. energy, without backst.
UMG 96RM-M Telegram Data point Discription DIF DIFE DIFE DIFE VIF VIFE 19 Uln - Phase L1 0x84 0x40 X X 0xFD 0x48 20 Uln - Phase L2 0x84 0x80 0x40 X 0xFD 0x48 21 Uln - Phase L3 0x84 0xC0 0x40 X 0xFD 0x48 22 I - Phase L1 0x84 0x40 X X 0xFD 0x59 23 I - Phase L2 0x84 0x80 0x40 X 0xFD 0x59 24 I - Phase L3 0x84 0xC0 0x40 X 0xFD 0x59 25 P - Phase L1 0x84 0x40 X X 0x2B X 26 P - Phase L2 0x84 0x80 0x40 X 0x2B X 27 P - Phase L3 0x84 0xC0 0x40
UMG 96RM-M M-Bus test Data string M-Bus $68$F7$F7$68$08$01$72$37$21$10$57$2E$28$09$02$02$00 $00$00$06$04$7E$18$00$00$00$00$86$10$04$7E$18$00$00 $00$00$86$20$04$00$00$00$00$00$00$86$40$04$28$00$00 $00$00$00$86$50$04$00$00$00$00$00$00$86$60$04$28$00 $00$00$00$00$86$80$40$04$92$18$00$00$00$00$84$40$24 $00$00$00$00$84$80$40$24$00$00$00$00$84$C0$40$24$00 $00$00$00$84$80$80$40$24$00$00$00$00$84$C0$80$40$24 $00$00$00$00$84$80$C0$40$24$00$00$00$00$04$24$FA$4F $00$00$84$80$80$40$FD$59$00$00$00$00$84$C0$80$40$2 B$00$
UMG 96RM-M Analysis via M-Bus Scanners (Excerpt) Data points 1 to 6 Note: The procedure for the verification of the M-Bus occurred with a M-Bus-Scanner of Wachendorff GmbH / Geisenheim. The figure shows a part of the software and is subject of the copy right of Wachendorff GmbH.
UMG 96RM-M Work values within the software GridVis Control of the values $187E = 6270 * 10 (resolution) = 62700 Wh 65
UMG 96RM-M Digital outputs The UMG 96RM-M has 2 digital outputs. The following functions can be optionally assigned to the digital outputs: Digital output 1 Address 200 = 0 Result of the comparator group 1 Address 200 = 1 Pulse output Digital output 2 Address 202 = 0 Result of the comparator group 2 Address 202 = 1 Pulse output Comparator group 1 State of digital output 1 Comparator B Logic Comparator A Result Addr. 616 Addr.
UMG 96RM-M Digital outputs - status indicators The status of the switching outputs is represented in the UMG 96RM-M display by circular symbols. C Since the display is only updated once per second, faster changes of the output states cannot be displayed. State of digital output 1 State of digital output 2 States of the digital output A current of <1 mA can flow. Digital output 1: Address 608 = 0 Digital output 2: Address 609 = 0 A current of <50 mA can flow.
UMG 96RM-M Pulse output Measured value selection (addr. 100, 101) Among other things, the digital outputs can also be used for the output of pulses to meter the energy consumption. After reaching a certain adjustable amount of energy, a pulse of defined length is applied to the output. Various adjustments must be made in order to use a digital output as a pulse output. Enter the power value here that is to be issued as an energy pulse. See Table 2.
UMG 96RM-M Pulse length (addr. 106) The pulse length applies for both pulse outputs and is permanently fixed via parameter address 106. Adjustment range: 1 .. 1000 1 = 10ms Default: 5 = 50ms Due to the minimum pulse length and minimum pulse pause, the values in the table are for the maximum number of pulses per hour. Pulse length Pulse pause Maximum pulses/hour The typical pulse length for S0 pulses is 30 ms.
UMG 96RM-M Pulse value (addr. 102, 104) The pulse value specifies how much energy (Wh or varh) should correspond to a pulse. The pulse value is determined by the maximum connected load and the maximum number of pulses per hour. If the pulse value is specified with a positive sign, pulses will only be issued if the measured value also has a positive sign. If the pulse value is specified with a negative sign, pulses will only be issued if the measured value also has a negative sign.
UMG 96RM-M Determining the pulse value External 230 V AC operating voltage Setting the pulse length Set the pulse length according to the requirements of the connected pulse receiver. For a pulse length of 30 ms, for example, the UMG 96RM can issue a maximum number of 60,000 pulses (see Table "Maximum Pulse Number") per hour.
UMG 96RM-M Limit value monitoring Two comparator groups are available for monitoring a limit value. Comparator group 1 is assigned to digital output 1 and comparator group 2 is assigned to digital output 2. State of digital output 1 UMG 96RM-M UMG 96RM Addr. 608 =0 Source selection Comparator group 1 Digital output 1 Inverter Comparator B Logic Comparator A Result Addr. 616 0/1 Addr. 200 =0 Comparator C Block diagram: Use of digital output 1 for limit value monitoring. 72 0/1 Adr.
UMG 96RM-M Example: Current monitoring in the neutral line If the current in the neutral line is greater than 100 A for 60 seconds, the digital output 1 should trip for at least 2 minutes. For the lead time of 60 seconds, any exceeding should be minimised. Address 112 = 60 seconds The following must be programmed: The operator for comparison between the measured value and the limit value. Address 113 = 0 (corresponds >=) 1. Comparator group 1 Select comparator group 1 for the limit value monitoring.
UMG 96RM-M Comparator Two comparator groups, each with 3 comparators, are available for monitoring limit values. The results from comparators A, B and C can be AND or OR linked. The linkage result from comparator group 1 can be assigned to digital output 1 and the linkage result from comparator group 2 is assigned to digital output 2. Comparator group 1 Comparator A Comparator B Comparator C Measured value (addr. 110) Limit value (addr. 108) Minimum turn-on time (addr. 111) Lead time (addr.
UMG 96RM-M • Measured value (addr. 110,116,122,129,135,141) The address of the measured value to be monitored is in the measured value. If measured value = 0, the comparator is inactive. • Limit value (addr. 108,114,120,127,133,139) Write the value in the limit that is to be compared with the measured value. • Minimum turn-on time (addr. 111,117,123,130,136,142) The linkage result (e.g. address 610) is maintained for the duration of the minimum turn-on time.
UMG 96RM-M Measured value Limit value Exceedance 2 seconds Lead time 2 seconds Minimum turn-on time Comparator result 76
UMG 96RM-M Parameter list comparator and digital outputs Address Format RD/WR Unit Note 100 SHORT RD/WR Address of the measured value, Digital output 1 101 SHORT RD/WR Address of the measured value, Digital output 2 102 FLOAT RD/WR Wh Pulse value, Digital output 1 104 FLOAT RD/WR Wh Pulse value, Digital output 2 106 SHORT RD/WR 10ms Minimum pulse length (1=10 ms) Digital output 1/2 107 SHORT RD/WR Result from comparator group 1; Link A, B, C (1=and, 0=or) 108 FL
UMG 96RM-M Address Format RD/WR Unit Note 122 SHORT RD/WR Comparator 1C, Address of the measured value 123 SHORT RD/WR s Comparator 1C, Minimum turn-on time 124 SHORT RD/WR s Comparator 1C, Lead time 125 SHORT RD/WR Comparator 1C, Operator „>=“=0 „<“=1 126 SHORT RD/WR Result from comparator group 2; Link A, B, C (1=and, 0=or) 127 FLOAT RD/WR Comparator 2A, Limit value 129 SHORT RD/WR Comparator 2A, Address of the measured value 130 SHORT RD/WR s Comparator 2A, Mi
UMG 96RM-M Address Note Adjustment Range Default 143 SHORT RD/WR s 144 SHORT RD/WR Comparator 2C, Lead time Comparator 2C, Operator „>=“ = 0 „<“ = 1 0..32000 0,1 0 0 200 SHORT RD/WR 201 SHORT RD/WR 202 SHORT RD/WR 203 SHORT RD/WR - Select the source for Digital output 1 Digital output 1 inverter Select the source for Digital output 2 Digital output 2 inverter 0..4 *1 0..1 *2 0 0 0..4 *1 0..
UMG 96RM-M Service and maintenance The device is subject to various safety tests prior to delivery and is marked with a seal. If a device is opened, the safety tests must be repeated. A warranty is only given for unopened devices. Service Repair and calibration We require the following information from you in order to deal with questions: Repairs and calibration can only be carried out by the manufacturer.
UMG 96RM-M 81
UMG 96RM-M Error messages The UMG 96RM-M shows three different error messages on the display: Symbol for an error message - warnings, - serious error and - metering range exceedances. Error cause Description of the error If there are warnings and serious errors, the error message is indicated by the symbol "EEE" followed by an error number. Symbol for an error message Example of error message 911: The error number is composed of serious error 910 and internal error cause 0x01.
UMG 96RM-M Warnings Internal causes of the error Warnings are less serious errors and can be acknowledged with buttons 1 or 2. The measured values continue to be recorded and displayed. This error is redisplayed after each voltage recovery. The UMG 96RM-M can usually determine the cause of an internal error and then report it with the following error code. The device must be sent to the manufacturer for inspection. Error EEE 500 Description of the error The mains frequency cannot be determined.
UMG 96RM-M Metering range exceedance Metering range exceedances are displayed for as long as they are present and cannot be acknowledged. A metering range is exceeded if at least one of the three voltage or current measuring inputs is outside of its specified metering range. The phase in which the metering range exceedance occurred is indicated with the "up" arrow. The "V" and "A" symbols show whether the metering range exceedance occurred in the current or voltage circuit.
UMG 96RM-M Parameters of the metering range exceedance A continuative error description is stored encoded in the parameters of the metering range exceedance (addr.
UMG 96RM-M Procedure in case of error Possibility of error Cause Help No display External fuse for the power supply has tripped. Replace fuse. No current display Measurement voltage not connected. Connect measurement voltage. Measurement current not connected. Connect measurement current. Current measurement in the wrong phase. Check and correct the connection if necessary. Voltage transformer factor incorrectly programmed.
UMG 96RM-M Possibility of error Cause Help Ind./cap. phase shift The current circuit is assigned to the wrong voltage circuit. Check and correct the connection if necessary. Real power is too small or too large. The programmed current transformer ratio is incorrect. Reading and programming the current transformer ratio on the current transformer The current circuit is assigned to the wrong voltage circuit. Check and correct the connection if necessary.
UMG 96RM-M Technical data General Net weight (with attached connectors) 300g Packaging weight (including accessories) 625g Device dimensions ca. l = 42mm, w = 97mm, h = 100mm Service life of the backlight 40,000 hours (50% of initial brightness) Transport and storage The following information applies for devices that are transported or stored in their original packaging.
UMG 96RM-M Supply voltage Installation overvoltage category 300V CAT II Protection of the power supply (fuse) 6 A, type C (approved by UL/IEC) Nominal range 20V - 250V (45..65Hz) oder DC 20V - 300V Working area +-10% from the nominal range Power consumption max. 4.5VA / 2W Connection capacity of the terminals (power supply) Connectable conductor. Only one conductor may be connected per contact point! Single-wire, multi-wire, finely stranded conductor 0.2 - 2.
UMG 96RM-M Connection capacity of the terminals (outputs) Rigid/flexible 0.14 - 1.5mm2, AWG 28-16 Flexible with ferrules without plastic sleeve 0.20 - 1.5mm2 Flexible with ferrules with plastic sleeve 0.20 - 1.5mm2 Tightening torque 0.20 - 0.
UMG 96RM-M Current measurement Rated current 5A Metering range 0 .. 6Arms Crest factor 1.98 Resolution 0.1 mA (display 0.01 A) Overvoltage category 300V CAT II Rated surge voltage 2kV Power consumption ca. 0.2 VA (Ri=5mOhm) Overload for 1 sec. 120 A (sinusoidal) Sampling rate 21.33 kHz (50 Hz), 25.6 kHz (60 Hz) per measuring channel Connection capacity of the terminals (voltage and current measurement) Connectable conductor.
UMG 96RM-M Serial interface M-Bus 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 baud M-Bus device load max. 20 mA Stripping length 7mm Connection capacity of the terminals (M-Bus) Single-wire, multi-wire, finely stranded conductor 0.20 - 1.5mm2 Pin terminals, ferrules 0.20 - 1.5mm2 Tightening torque 0.20 - 0.
UMG 96RM-M 93
UMG 96RM-M Parameters of functions Function Symbol Accuracy class Metering range Display range Total real power P 0.55) (IEC61557-12) 0 .. 5.4 kW 0 W .. 999 GW * Total reactive power QA, Qv 1 (IEC61557-12) 0 .. 5.4 kvar 0 varh .. 999 Gvar * Total apparent power SA, Sv 0.55) (IEC61557-12) 0 .. 5.4 kVA 0 VA .. 999 GVA * Total active energy Ea 0.55) (IEC61557-12) 0 .. 5.4 kWh 0 Wh .. 999 GWh * Total reactive energy ErA, ErV 1 (IEC61557-12) 0 .. 5.4 kvarh 0 varh ..
UMG 96RM-M Function Symbol Accuracy class Metering range Display range Current harmonics Ih Class 1 (IEC61000-4-7) up to 2.5 kHz 0 A .. 999 kA THD of the current 3) THDi 1.0 up to 2.5 kHz 0 % .. 999 % THD of the current 4) THD-Ri - - - Mains signal voltage MSV - - - 1) Referred to amplitude. 2) Referred to phase and amplitude. (IEC61557-12) * The display returns to 0 W when the maximum total energy values are reached. 5) Accuracy class 0.5 with ../5 A transformer.
UMG 96RM-M Parameter and Modbus address list The following excerpt from the parameter list contains settings that are necessary for proper operation of the UMG 96RM, such as current transformers and device addresses. The values in the parameter list can be written and read. In the excerpt, the measured value list files the measured and calculated measured values, output status data and recorded values so that they can be read.
UMG 96RM-M Adress Format RD/WR Unit 24 FLOAT RD/WR V 26 FLOAT RD/WR A 28 FLOAT RD/WR A 30 FLOAT RD/WR V 32 FLOAT RD/WR V 34 SHORT RD/WR Hz 35 SHORT RD/WR 36 SHORT RD/WR Note Adjustment Range Default Voltage transformer V2, sec. Current transformer I3, primary Current transformer I3, sec. Voltage transformer V3, prim. Voltage transformer V3, sec. Frequency determination 0=Auto, 45 .. 65=Hz Display contrast 0 (low), 9 (high) Backlight 0 (low), 9 (high) 100, 400 0..1000000 1..5 0..
UMG 96RM-M Adress Format RD/WR Unit 81 SHORT RD/WR 82 SHORT RD/WR 83 SHORT RD/WR 84 SHORT RD/WR 500 SHORT RD/WR 501 SHORT RD/WR 502 SHORT RD/WR 503 SHORT RD/WR 504 SHORT RD/WR 505 SHORT RD/WR 506 SHORT RD/WR 507 SHORT RD/WR 508 SHORT RD/WR - Note Adjustment Range Secondary address, extended section 1 Secondary address, extended section 2 Secondary address, extended section 3 Secondary address, extended section 4 Terminal assignment, I L1 Terminal assignment, I L2 Terminal assi
UMG 96RM-M C Only the first three positions (###) of a value are shown on the display. Values larger than 1,000 are marked with „k”.
UMG 96RM-M Addresse 840 842 844 846 848 850 852 854 856 858 860 862 864 866 868 870 872 874 876 878 880 882 884 886 888 890 892 894 896 898 900 902 100 Format RD/WR Unit Note FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD % % % % V V V V V V A A A A W W W W var v
UMG 96RM-M Addresse 904 906 908 910 912 914 916 918 920 922 924 926 928 930 932 934 936 938 Format RD/WR Unit Note FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT FLOAT RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD RD var var % % % % % % A A A A A A - Harmonic distortion power D3 L3N Sum; Dsum3=D1+D2+D3 THD1 I1, based on I01 THD2 I2, based on I02 THD3 I3, based on I03 TDD1 I1, based on rated load current TDD2 I2, based on rated l
UMG 96RM-M Dimensional drawings All dimensions in mm. Rear view Side view 42 102 100 □ 91,5 □ 96 max.
UMG 96RM-M Bottom view Cutout dimensions 92+0,8 92+0,8 □ 91,5 97 103
UMG 96RM-M Overview of measured value displays with information regarding the associated profile display A01 (Profile: 1,2,3) B01 (Profile: 1,2,3) C01 (Profile: 1,2,3) D01 (Profile: 1,2,3) Measured values L1-N voltage L2-N voltage L3-N voltage Mean values L1-N voltage L2-N voltage L3-N voltage Maximum values L1-N voltage L2-N voltage L3-N voltage Minimum values L1-N voltage L2-N voltage L3-N voltage A02 (Profile: 1,2,3) B02 (Profile: 1,2,3) C02 (Profile: 1,2,3) Measured values L1-L2 voltage L2-L
UMG 96RM-M A06 (Profile: 1,2,3) B06 (Profile: 1,2,3) C06 (Profile: 1,2,3) D06 (Profile: 1,2,3) Measured value Sum Active power Mean value Sum Active power Maximum value Sum Active power Maximum value Sum Active power mean value A07 (Profile: 1,2,3) B07 (Profile: 1,2,3) C07 (Profile: 1,2,3) Measured values L1 apparent power L2 apparent power L3 apparent power Mean values L1 apparent power L2 apparent power L3 apparent power Maximum values L1 apparent power L2 apparent power L3 apparent power A
UMG 96RM-M A12 (Profile: 1,2,3) B12 (Profile: 1,2,3) C12 (Profile: 1,2,3) Measured value Distortion factor (THD) I L1 Measured value Distortion factor (THD) I L2 Measured value Distortion factor (THD) I L3 A13 (Profile: 1,2,3) B13 (Profiel: 1,2,3) C13 (Profile: 1,2,3) Maximum value Distortion factor (THD) U L1 Maximum value Distortion factor (THD) U L2 Maximum value Distortion factor (THD) U L3 A14 (Profile: 1,2,3) Maximum value Distortion factor (THD) I L1 B14 (Profile: 1,2,3) Maximum value
UMG 96RM-M A18 (Profile: 1,2,3) B18 (Profile: 1,2,3) C18 (Profile: 1,2,3) D18 (Profile: 1,2,3) Measured value Total active energy Measured value Total active energy Measured value Total active energy Measured value Sum Apparent energy (without a backstop) (import) (export) A19 (Profile: 1,2,3) B19 (Profile: 1,2,3) C19 (Profile: 1,2,3) Measured value (ind) Reactive energy Measured value Sum Reactive energy cap. Measured value Sum Reactive energy ind.
UMG 96RM-M A24 (Profile: 2) B24 (Profile: 2,3) Measured value 1st. harmonic I L1 Measured value 3rd. harmonic I L1 G24 (Profile: 2,3) H24 (Profile: 2) ... Measured value 13th. harmonic I L1 Measured value 15th. harmonic I L1 ... A25 (Profile: 2) B25 (Profile: 2,3) G25 (Profile: 2,3) H25 (Profile: 2) Measured value 1st. harmonic I L2 Measured value 3rd. harmonic I L2 Measured value 13th. harmonic I L2 Measured value 15th.
UMG 96RM-M A30 (Profile: 2) B30 (Profile: 2) Maximum value 1st. harmonic I L1 Maximum value 3rd. harmonic I L1 G30 (Profile: 2) H30 (Profile: 2) ... Maximum value 13th. harmonic I L1 Maximum value 15th. harmonic I L1 ... A31 (Profile: 2) B31 (Profile: 2) G31 (Profile: 2) H31 (Profile: 2) Maximum value 1st. harmonic I L2 Maximum value 3rd. harmonic I L2 Maximum value 13th. harmonic I L2 Maximum value 15th. harmonic I L2 A32 (Profile: 2) Maximum value 1st.
UMG 96RM-M Declaration of conformity The product fulfils the following EC Directives: 2004/108/EG Electromagnetic compatibility of electrical equipment. 2006/95/EG Electrical equipment for use within certain voltage limits.
UMG 96RM-M Connection example = MM+ + UMG 96RM-M MM+ M-Bus Digital outputs Supply voltage Voltage input Current input N/- L/+ V1 V2 V3 VN 1 3 4 5 6 2 S2 S1 S2 S1 S2 Consumer S2 S2 S2 230V/400V 50Hz S1 S1 S1 S1 L1 L2 1) 2) L3 N 3) UL/IEC approved overcurrent protection device (6 A, type C) UL/IEC approved overcurrent protection device (10 A, type C) Jumpers (external) 111
UMG 96RM-M Brief instructions Current transformer setting, primary current Changing the current transformer setting Display Programming mode Switch to programming mode: • Simultaneously press buttons 1 and 2 for approximately 1 second in order to switch to programming mode. The symbols for the programming mode (PRG) and for the current transformer (CT) appear. • Confirm the selection with button 1. • The first digit of the input area for the primary current starts flashing.