Datasheet

12
Scope of operation and types of variants
General technical data
UMG 511 UMG 605
Supply voltage L-N, AC Refer to product variant overview Refer to product variant overview
Overvoltage category 500V CATIII 300V CATIII
Quadrants 4 4
Continuous measurement yes yes
8 channel scanning rate Per channel 20 kHz 20 kHz
Weight 1kg 350g
Dimensions L=144mm * B=144mm * H=81 mm L=107,5mm*B=90mm*H=76/82mm
Mounting According to IEC EN 60999-1/DIN EN 50022 Front panel mounting 35mm DIN rail
Working temperature range -10…50 °C -10…55 °C
Connectable conductor (U/I) Single wire, multi-wire, fine-wire 0,08 - 2,5 mm² 0,08 - 2,5 mm²
pin cable lugs, ferrule 1,5 mm² 1,5 mm²
Protection class According to EN 60529 IP 50 front /IP 20 rear IP 20
- = not possible = contained
Overview of product variants UMG 511 / UMG 605
Three/four phase power quality analysers; current transformer .../1/5a; including GridVis programming and analysis software
95...265V AC,
100...370V DC
123...240V AC,
175...340V DC
50...130V AC,
70...180V DC
20...50V AC,
20...70V DC
4 voltage and
4 current inputs
Memory
128/256 MB Flash
digital inputs
digital outputs
1 temperature input
RS 232
RS 485
Ethernet 100baseT
Profibus DP V0
Type
Item number
8 5 - -
U
MG 511
5
2.19.001
2 2
U
MG 605
5
2.16.027
2 2
UMG 605 52.16.028
2 2
U
MG 605
5
2.16.029
O
ptions (for all versions)
Emax function application program (peak demand management) Emax 52.16.080
BACnet communication BACnet 52.16.081
Supply voltage
I
nterfaces
Measurement range
UMG 511 UMG 605
L-N voltage, AC (without voltage transformer) Free voltage transformer settings 5 …500 VAC 5…500 VAC
L-L voltage, AC (without voltage transformer) Free voltage transformer settings 8…870 VAC 8…870 VAC
Current (transformer: x/1 and x/5A) 0,005..6 A 0,005..6 A
Frequency of mains 40 ..70 Hz 40 ..70 Hz
Networks TN, TT IT, TN, TT
Measurement in single/multi-phase networks
1 ph, 2 ph, 3 ph, 4 ph
and up to 4 x 1 ph
1 ph, 2 ph, 3 ph, 4 ph
and up to 4 x 1 ph
Periphery
UMG 511 UMG 605
Digital inputs Status, logic or pulse input 8 2
Digital outputs Switch logic output or pulse output 5 2
Temperature measurement input PT100, PT1000, KTY83, KTY84 - 1
Password protection Multilevel yes yes
Demand management Optional 64 channels ja ja
Software GridVis ja ja
Features
UMG 511 UMG 605
Memory 256 MB 128 MB
Clock +/- 1 min per month +/- 1 min per month
Integrated logic Programming language Jasic
®
Programming language Jasic
®
Operating hour meter yes yes
Weekly time switch Jasic
®
Jasic
®