Setup Guide
Table Of Contents
- Intelis Gas Meter Proving and Setup Guide
- Contents
- New in this Document
- 1 Important Safety and Compliance Information
- USA, FCC Part 15 Spectrum Compliance
- Canada, ISED Spectrum Compliance
- RF Exposure (FCC/ISED)
- Australia, ACMA Spectrum Compliance
- Electromagnetic Compatibility
- Intrinsic Safety
- Lithium Battery
- Transportation Classification
- Electrostatic Discharge
- Electrostatic Ignition Hazard
- Device Cleaning
- Do Not Drop
- Product Standards
- 2 Overview
- 3 Proving
- 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System
- 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System
- 6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System
- 7 Calibrating the Intelis Gas Meter
4. Select Finish to return to the Tools screen.
To Unseal an Intelis Ggas Meter's UMU
1. From the Tools menu, select OWR Intelis Gas Riva.
2. Enter the Endpoint ID.
3. Select the Endpoint Mode.
4. Select Next.
5. Select Test Mode Operations.
6. Select the Utility ID.
7. Click Next.
8. Select Adjust Calibration for a device with a Seal State of Sealed. A warning message
appears indicating that Adjusting Calibration will unseal the meter.
9. Select Yes.
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 65 of 66