Setup Guide

Valid values:
Cubic Meters Per Hour. 0.01-11.32
Cubic Feet Per Hour. 0.01 to 400.00
Station Mode of % Error. -7.00% to +7.00%
Station Mode of % Proof. 93.00% to 107.00%
Station Mode of % Accuracy. 93.00% to 107.00%
Note: The Prover flow rate check value must be lesser than the Prover flow rate open
value.
Prover percentage check. Enter the value from the test mode screen for the meter's check
result. (Example entry +0.7).
Prover flow rate open. Enter the prover station open flow rate. This is typically 80-100% of
the capacity. (Example entry 250).
Note: The Prover flow rate open value must be greater than the Prover flow rate check
value.
Prover percentage open. Enter the value from the prover screen for the meter's open
result. (Example entry -0.3).
8. Select Next. The Adjust Calibration Summary appears so that you may review your
parameter changes.
9. Select Next. A progress bar indicates that the meter's calibration is being adjusted.
The meter uses the entered values to calculate the required calibration parameters to a
target of one of the following, depending on your selected prover station mode:
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 62 of 66
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter