14. Verify or enter Flow rates (Air Rate): Open, Check, Other settings. Itron recommended
typical settings are listed. For metric flow rates: choose from the flow rate dropdown list
options. These convert the Imperial (CFH) flow rates to Metric (CMH). The Other flow rate is
not required if the Test Revs: Index-Other is set to 0.
–
Open: 250
–
Check: 50
–
Other: 125
Important! Required Open and Check flow rates may be different than those
shown here.
15. Select the Proving Method: Index. The options are Index or Magnetic.
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 40 of 66
Intelis Gas Meter Proving and Setup Guide
5 Proving the Intelis Gas Meter Using the Measurement Systems
Proving System