Note: Testing has shown acceptable results with settings as low as 4 pulses on open
and 1 on check but Itron recommends more pulses to obtain better results.
13. Select the Volume: Index setting. The Volume: Index setting is the recommended volume
per pulse. Itron recommended settings are listed.
–
Imperial setting: 0.25 cu ft/rev
–
Metric setting: 0.01 m3/rev
Note: Intelis Gas Meter volume per pulse is in dm3; 10dm3 = 0.01 m3.
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 39 of 66
Intelis Gas Meter Proving and Setup Guide
5 Proving the Intelis Gas Meter Using the Measurement Systems
Proving System