Tip: The Measurement Systems application software on the Measurement Systems
proving system provides touch screen functionality. You may prove the Intelis Gas
Meter using the Measurement Systems Prover screen or a traditional mouse-click to
complete the steps.
Itron recommends these Intelis Gas Meter Test Mode parameter settings for the Measurement
Systems proving system:
■
Test Mode pulse weight: 0.25 CF
■
Test Mode pulse width: 150 ms
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 32 of 66
Intelis Gas Meter Proving and Setup Guide
5 Proving the Intelis Gas Meter Using the Measurement Systems
Proving System