Setup Guide
Table Of Contents
- Intelis Gas Meter Proving and Setup Guide
- Contents
- New in this Document
- 1 Important Safety and Compliance Information
- USA, FCC Part 15 Spectrum Compliance
- Canada, ISED Spectrum Compliance
- RF Exposure (FCC/ISED)
- Australia, ACMA Spectrum Compliance
- Electromagnetic Compatibility
- Intrinsic Safety
- Lithium Battery
- Transportation Classification
- Electrostatic Discharge
- Electrostatic Ignition Hazard
- Device Cleaning
- Do Not Drop
- Product Standards
- 2 Overview
- 3 Proving
- 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System
- 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System
- 6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System
- 7 Calibrating the Intelis Gas Meter
Test Mode is used in a meter shop or by a regulatory body. Some meter configuration and prover
setup are required for the prover to properly interpret the meter pulses. Meter configuration is
completed using FDM.
Proving accessory Itron part number
Honeywell SNAP pulse output prover cable CFG-7100-100
Measurement Systems pulse prover cable CFG-7100-500
Energy Economics (EEI) pulse prover cable CFG-7100-400
Can-Tronics pulse prover cable CFG-7100-200
Itron BPG SONICAL SN G6 pulse prover cable CFG-7100-700
Intelis Gas Meter proving stand FIX-7100-001
Table 3Accessory name and part number
You must exit Test Mode to program Test Mode parameters. Attempting to program Test Mode
parameters while the meter is in Test Mode will result in a temporary error. The configurable
parameters for Test Mode are listed in the table below.
Parameter Default Range
Pulse width 150 ms. 10 ms–1000ms (1 second)
Pulse weight 0.25 cf/pulse
■
0.10CF
■
0.25CF
■
0.50CF
■
1.0CF
■
10dm3
■
50dm3
Maximum Test Mode time 1 hour
Important:The Intelis Gas Meter will
automatically time out after 1 hour in
Test Mode whether there is activity or
not.
1–255 hours (whole numbers only)
Table 4Test Mode configurable parameters
Test Mode configuration can be completed at the factory at the time the meter is manufactured.
Parameters are configured in one of three ways:
■
Through a custom programming file.
■
Through programmed default configuration.
■
In the field using FDM Tools.
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 16 of 66
Intelis Gas Meter Proving and Setup Guide 3 Proving