Intelis Gas Meter Proving and Setup Guide REFERENCE GUIDE 08 July 2020 knowledge to shape your future 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide Intelis Gas Meter Proving and Setup Guide 08 July 2020 815-0154-00 REV 000 Copyright © 2020 Itron, Inc. All rights reserved. Confidentiality Notice Confidential Information of Itron®, Inc., provided under nondisclosure obligations.
Contents In te lsi Ga s Me te rP o r vn i g a n d Se tu p Gu d i e 1 Co n te n ts 3 Ne win th si Do c u me n t 5 1 2 3 Important Safety and Compliance Information 6 USA, FCC Part 15 Spectrum Compliance 6 USA, FCC Class B-Part 15 6 Modifications and Repairs 6 Canada, ISED Spectrum Compliance 7 RF Exposure (FCC/ISED) 7 Australia, ACMA Spectrum Compliance 7 Electromagnetic Compatibility 7 Intrinsic Safety 7 Lithium Battery 8 Transportation Classification 8 Electrostatic Discharge
Intelis Gas Meter Proving and Setup Guide 4 5 6 7 08 July 2020 Table of Contents Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System 19 Tempering the Meter by Exercising 20 Setting Up the Prover 21 Proving the Intelis Gas Meter 26 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 31 Temper the Meter by Exercising the Meter 33 Setting Up the Measurement Systems Prover 34 Proving the Intelis Gas Meter 45 Proving the Intelis Gas Meter Using the En
Intelis Gas Meter Proving and Setup Guide Table of Contents New in this Document Revision Date Description REV 000 July 2020 Original publication date. Note: The content in this document used to be part of the Intelis Gas Meter Installation Guide, which has since been redacted and merged with the Intelis Gas Meter Technical Reference Guide. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
1 Important Safety and Compliance Information This section provides important information for your safety and product compliance. USA, FCC Part 15 Spectrum Compliance This device complies with Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. Operation is subject to the following two conditions: ■ This device may not cause harmful interference.
Intelis Gas Meter Proving and Setup Guide 1 Important Safety and Compliance Information Canada, ISED Spectrum Compliance Compliance Statement Canada Déclaration de Conformité This device complies with Innovation, Science and Economic Development Canada (ISED) license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Intelis Gas Meter Proving and Setup Guide 1 Important Safety and Compliance Information Lithium Battery Warning! Follow these procedures to avoid injury to yourself or others. ■ The lithium battery may cause a fire or chemical burn if it is not disposed of properly. ■ Do not recharge, disassemble, heat above 212°Fahrenheit (100°C Celsius), crush, expose to water, or incinerate the lithium battery. Fire, explosion, and severe burn hazard.
Intelis Gas Meter Proving and Setup Guide 1 Important Safety and Compliance Information Product Standards These instructions are not intended to replace any utility or company-established meter installation procedures. These instructions are provided for additional information. The meter installation must comply with all country, state, and local building and safety regulations as well as federal regulations including Section 192.353 of Title 49 of the Code of Federal Regulations.
2 Overview The Intelis Gas Meter is a solid state ultrasonic smart meter equipped with the following features: ■ High flow alarm with optional automatic valve shutoff ■ High temperature alarm with an optional automatic valve shutoff ■ Built in temperature conversion ■ Integrated RF communications module ■ Air detection ■ Reverse flow detection The internal automatic safety shutoff valve can be configured to function similarly to an excess flow valve downstream of the meter.
Intelis Gas Meter Proving and Setup Guide 2 Overview Caution: Always follow your company's procedures for gas meter installation and operation. Related Documents For more information about the Intelis Gas Meter and compatible system architecture documentation, see your product's documentation. The following table lists documentation that might be relevant to the setup of your meter. All Itron product documentation can be located and accessed through https://access.itron.com/.
Intelis Gas Meter Proving and Setup Guide 2 Overview Table 1 Intelis Gas Meter specifications (continued) Specification Specification value ■ 20LT ■ 30LT ■ 1A/Sprague standard ■ 1-1/4", 1" Pittsburgh ■ ISO G 1-1/4" Meter MAOP 5 PSIG (35 kPa) Valve Actuated swing valve, maximum 0.035 CFH (1 L/H) leakage rate LCD displayed units CCF (100 x cubic foot) or cubic meter LCD resolution 00000.001 CCF (0.1 CF) or 00000.001 m3 Case Aluminum case with ASA 49 gray powder coat finish Weight 4.5 lbs (2.
Intelis Gas Meter Proving and Setup Guide 2 Overview Table 1 Intelis Gas Meter specifications (continued) Specification Specification value ■ Valve: -13° F (-25° C) to +131° F (55° C) ■ RF: -40° F (-40° C) to +158° F (70° C) ■ LCD: -40F (-40C) to 185F (85C)1 Storage temperature ratings -40° F (-40° C) to +158° F (70° C) Battery information 4 Lithium manganese dioxide (LiMnO2) 'A' cell batteries, replaceable Battery life 20 years for meter, RF, and valve using recommended operating parameters Badgi
Intelis Gas Meter Proving and Setup Guide 2 Overview Table 2 Intelis Gas Meter dimensions A B C D E F G 8.4 inches 6.4 inches 6.0 inches 5.4 inches 3.1 inches 9.1 inches 1.9 inches 214.4 mm 161.9 mm 152.4 mm 138.3 mm 78.6 mm 230.6 mm 48.8 mm Note: B dimension can vary with different hubs options. (20LT is shown). 08 July 2020 815-0154-00 REV 000 Itron, Inc.
3 Proving This section provides information about proving the Intelis Gas Meter. Meteorological proving allows users to test the Intelis Gas Meter accuracy. Temper meters unpackaged in the proving area for 24 hours prior to the testing. Warning! Follow your company's standard operating procedures regarding the use of personal protection equipment (PPE). Adhere to guidelines issued by your company in addition to those contained in this document when proving meters.
Intelis Gas Meter Proving and Setup Guide 3 Proving Test Mode is used in a meter shop or by a regulatory body. Some meter configuration and prover setup are required for the prover to properly interpret the meter pulses. Meter configuration is completed using FDM.
Intelis Gas Meter Proving and Setup Guide 3 Proving Note: Refer to the instructions for your prover type in the following sections for the recommended Test Mode pulse width and pulse weight parameters. Entering Test Mode Entering Test Mode is accomplished using either the FDM mobile client or through the magnet in the prover connection end of the prover cable. After the Intelis enters Test Mode, the following steps occur. 1. The Intelis Gas Meter logs an Entering Test Mode event. 2.
Intelis Gas Meter Proving and Setup Guide 3 Proving Prover Types The Intelis Gas Meter compatible prover types are listed below. See the procedure for your prover type. For questions about your specific prover type, see the manufacturer's prover documentation. 08 July 2020 ■ Honeywell SNAP ■ Measurement Systems ■ Energy Economics, Inc (EEI) Sonic Nine ■ Can-Tronics (TBD) ■ Itron BPG SONICAL SNG6 (TBD) 815-0154-00 REV 000 Itron, Inc.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System This section provides the information to set up an Intelis Gas Meter using a Honeywell SNAP proving system. Note: Proving the Intelis Gas Meter requires Honeywell SNAP prover software .NET X.Y. Temper meters unpackaged in the proving area for 24 hours prior to testing. If tempering is not possible,follow the instructions described in Tempering the Meter by Exercising on page 20.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Intelis Gas Meter Proving and Setup Guide Itron recommends these Intelis Gas Meter Test Mode parameter settings when testing on the Honeywell SNAP proving system: ■ Test Mode pulse weight: 0.25 CF ■ Test Mode pulse width: 150 ms Tempering the Meter by Exercising 1. Log on to the prover as a Supervisor. 2. Select Special Functions. 3. Select Diagnostics (local). 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System 4. After the Diagnostics screen opens, clamp the meter in place by pressing Clamp in the lower right screen corner. 5. Press the clamp buttons on both sides of the SNAP prover. 6. After the Intelis Gas Meter is clamped to the prover, turn on the appropriate nozzles in the center of the screen to exercise the meter. Note: Typically, Nozzle Number 2-245 CFH is used to exercise the Intelis Meter.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Intelis Gas Meter Proving and Setup Guide 3. Select Edit Config Tables. 4. Select Edit Setup Values. 5. Select the MeterMfg_Table checkbox. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Intelis Gas Meter Proving and Setup Guide 6. Click in the next available (open) row in the top left table and enter ITRON-RTY for the meter manufacturer. Click off of the table and select Update Table. Important! If Itron, Actaris, Sprague, or other is listed, it is still necessary to add a new manufacturer name specifically for the Intelis Gas Meter. Itron suggests using ITRON-RTY following the steps listed below.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Intelis Gas Meter Proving and Setup Guide This step allows you to copy an existing meter setup and modify it to create the new Intelis Gas Meter (for example, choose I-250, Itron). Select a similar meter's row. Follow the MeterTable instructions: Click on the next open/empty row. To complete the addition, click off of the table and select Add Chosen Meter.
Intelis Gas Meter Proving and Setup Guide 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System – Check revs: 2 – Other revs: 1 – Exercise revs: 1 – Revs unit volume: 4 – Jog rate: 50 – Max press diff: 0.5 – Miscellaneous utility-specific parameters as required (covered in Step 9) 9. Verify or enter the TOLERANCE values for POS, NEG, Slope, Repair, and Meter Size. These settings (values) are utility-specific. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Intelis Gas Meter Proving and Setup Guide 10. Click off the table and select Update Table. 11. Exit the program. Proving the Intelis Gas Meter 1. Select Prove Meter. 2. Select the meter manufacturer (ITRON-RTY). 3. Select Intelis 250. 4. Select the correct mode for your meter (TC or NTC). Note: REG is equivalent to NTC 5. Select the meter results mode (% error, % accuracy, % proof). 08 July 2020 815-0154-00 REV 000 Itron, Inc.
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Intelis Gas Meter Proving and Setup Guide 6. Select the test mode (intest, outtest). 7. Select the optic test method (optic, mag). 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System 8. Select the correct Index (volume per pulse) setting (Imperial, ¼ CF; Metric, 10 dm3). 9. Click Select Meter to advance to the next screen. 10. Connect the SNAP pulse prover cable end (Itron part number CFG-7100-100) to the J4 (1) 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System optical sensor port typically on the top of the SNAP prover electronic enclosure, next to the meter temperature probe port. 11. Connect the Intelis Gas Meter end of the SNAP pulse prover cable to the Intelis Gas Meter optical port by firmly pressing the aluminum round threads of the cable into the port. Begin by positioning the cable at the 4:30 o'clock position.
Intelis Gas Meter Proving and Setup Guide 4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving System Caution: Do not over tighten this, as doing so could damage the index housing or the proving cable head. Note: The Intelis Gas Meter will automatically go into Test Mode (Menu 6) when the pulse prover output cable is connected. 12. Start the proving test by clicking RUN on the SNAP prover. 13. Verify the LED blinks (pulses) when the air begins to flow through the meter.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System This section provides the information to set up an Intelis Gas Meter using a Measurement Systems proving system. Note: Proving the Intelis Gas Meter requires Measurement Systems prover software. See the Measurement Systems prover documentation for version requirements. Temper meters unpackaged and uncapped in the proving area for 24 hours prior to testing.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide Tip: The Measurement Systems application software on the Measurement Systems proving system provides touch screen functionality. You may prove the Intelis Gas Meter using the Measurement Systems Prover screen or a traditional mouse-click to complete the steps.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide Temper the Meter by Exercising the Meter 1. Open the Measurement Systems Proving System application software. 2. Select Meter Setup to exercise the meter. 3. Select the Global tab > Operate Valves and View Sensor Readings. 4. From the Operate Valves and Observe Sensor Readings screen, press the 1 key or Chuck Valve to clamp the meter in place.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide 5. After the Intelis Gas Meter is clamped to the prover, turn on the appropriate nozzle in the center of the screen to exercise the meter. Note: Nozzle 6-250 is typically used for the Intelis Gas Meter. 6. Exercise the meter for 10 to 15 minutes before proving the meter. 7. Close the nozzle. 8. Unclamp the meter by pressing the Chuck Valve button. 9.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide 3. Select the Meter Model-1 tab. 4. Select the meter model. If the Intelis 250 or Intelis 250 TC are set up in the MMS prover, go to step 10. If the Intelis 250 or Intelis 250 TC is not listed, continue to step 5. 5. Select the Prover tab. Note: To access the Prover tab, you will be asked for a password. Contact Measurement Systems if you do not have a password.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide 6. Select Enter New Meter Model. 7. Select an existing meter model similar to the new meter as a template. Itron recommends using the S-250 for an Intelis 250 gas meter and an S-250 TC for the Intelis 250 TC gas meter. Select the green bar on the bottom of the screen to use the selected meter as the starting point.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 8. Enter the new meter name in the New Meter Model Name text box (1) and click OK (2). 9. Press Enter or click the blue bar at the bottom of the screen to go back to the meter setup menu. 10. Select the Exercise Cycle Duration. The Intelis Gas Meter Exercise Cycle Duration is typically 15 seconds. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 11. Set Leak Test to Yes. 12. Verify or enter Test Revs: Index Open, Check, Other settings. Itron recommended settings are listed. 08 July 2020 – Open: 6 – Check: 2 – Other: 0 815-0154-00 REV 000 Itron, Inc.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide Note: Testing has shown acceptable results with settings as low as 4 pulses on open and 1 on check but Itron recommends more pulses to obtain better results. 13. Select the Volume: Index setting. The Volume: Index setting is the recommended volume per pulse. Itron recommended settings are listed. – Imperial setting: 0.25 cu ft/rev – Metric setting: 0.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 14. Verify or enter Flow rates (Air Rate): Open, Check, Other settings. Itron recommended typical settings are listed. For metric flow rates: choose from the flow rate dropdown list options. These convert the Imperial (CFH) flow rates to Metric (CMH). The Other flow rate is not required if the Test Revs: Index-Other is set to 0.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide 16. Select the mode for your meter. Available options are TC or NON-TC. 17. Select the Meter Model-2 tab. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 18. Verify or enter the InTest and OutTest limit values for Open Limits, Check Limits, Other Limits, and Delta Limit. The Open, Check, Other, and Delta Limit settings (values) are utilityspecific. 19. Verify or enter the limit values for Check Dif Limits, Monitor Peaks?, and MaxPeak. The Check Dif Limits, Monitor Peaks?, and MaxPeak settings (values) are utility-specific.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 20. Verify or enter Tangent Adjustment: Tangent Type. Note: The Tangent Adjustment: Tangent Type setting can be any non-blank value. The Tangent Adjustment: Tangent Type value is not used to prove the Intelis Gas Meter. 21. Select the Global tab and Display Method for your meter results. Display Method options are listed.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 22. Select InTest/OutTest. The options are InTest or OutTest. 23. Press or click the green bar at the bottom of the screen to exit the Meter Setup Menu. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide Proving the Intelis Gas Meter 1. Connect the Measurement Systems pulse prover cable (Itron part number CFG-7100-500) to the port on the left side of the Measurement Systems prover. Note: If your Measurement Systems prover does not look like the illustration, contact Measurement Systems for an adapter cable to connect the Measurement Systems pulse prover cable to your prover.
5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System Intelis Gas Meter Proving and Setup Guide 2. Connect the Intelis Gas Meter end of the Measurement Systems pulse prover cable to the Intelis Gas Meter by firmly pressing the aluminum round threads of the cable into the port. Begin by positioning the cable at the 4:30 o'clock position. Tighten the cable to the 6:30 o'clock position.
Intelis Gas Meter Proving and Setup Guide 5 Proving the Intelis Gas Meter Using the Measurement Systems Proving System 4. Press or click the green bar on the bottom of the screen to begin the Measurement Systems proving test. 5. Verify the LED blinks (pulses) when the air begins to flow through the meter. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System This section provides the information to set up an Intelis Gas Meter using an EEI proving system. Note: Proving the Intelis Gas Meter requires Energy Economics prover software. Temper meters unpackaged in the proving area for 24 hours prior to testing. If tempering is not possible, follow the instructions described in Temper the Meter by Exercising the Meter on page 50.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Intelis Gas Meter Proving and Setup Guide Itron recommends these Intelis Gas Meter Test Mode parameter settings when testing onthe EEI proving system: 08 July 2020 ■ Test Mode pulse weight: 0.10 CF ■ Test Mode pulse width: 150 ms 815-0154-00 REV 000 Itron, Inc.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Intelis Gas Meter Proving and Setup Guide Temper the Meter by Exercising the Meter 1. Log on to the EEI prover application. Note: Logging on to the EEI prover application requires maintenance level security. 2. Select Maintain > Sensor Maintenance. 3. Clamp the Intelis Gas Meter to the prover and turn on appropriate nozzles in the center column of the Sensor Maintenance window to exercise the meter.
Intelis Gas Meter Proving and Setup Guide 6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Note: Nozzle 192 CFH is typically used to exercise the Intelis Gas Meter. 4. Exercise the meter for 10 to 60 minutes until the Out Temp value does not vary by more than 0.2° F within a minute. 5. Close the nozzle. 6. Unclamp the meter. 7. Exit the Sensor Maintenance window by clicking Done in the lower right corner of the window. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Intelis Gas Meter Proving and Setup Guide Setting Up the EEI Prover 1. Log on to the EEI prover application. 2. Clamp the Intelis Gas Meter to the EEI prover. 3. Select the Meter tab and choose ITRON from the Manuf drop-down list. Important! If ITRON is not in the Manuf. drop-down list, contact EEI Prover Systems for help in adding ITRON to your EEI Systems Prover. 4.
Intelis Gas Meter Proving and Setup Guide 6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System 8. Select Open & Check for the Run Type Test Options. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Intelis Gas Meter Proving and Setup Guide 9. Enter 100 for Flow > Open %. Important! Required Open flow rates may be different than those shown here. 10. Enter 20 for Flow > Check %. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Intelis Gas Meter Proving and Setup Guide Important! Required Check flow rates may be different than those shown here. 11. Select 1/10 cuft for Hand Size. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System 12. Click Done to complete the Meter Information > Test Options and close the window. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System Intelis Gas Meter Proving and Setup Guide Note: All other settings (Tangents, Ultra Intest Lock, Ultra Adjust Lock and Station) on the Test Options window are not relevant for the Intelis meter. Proving the Intelis Gas Meter 1. Connect the EEI pulse prover cable (Itron part number CFG-7100-400) to the EEI prover index eye connector. 2.
Intelis Gas Meter Proving and Setup Guide 6 Proving the Intelis Gas Meter Using the Energy Economics (EEI) Proving System 3. Press the EEI prover's green button to begin the proving test. 4. Verify the LED blinks (pulses) when the air begins to flow through the meter. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
7 Calibrating the Intelis Gas Meter FDM allows users to re-calibrate the Intelis Gas Meter accuracy. The Intelis Gas Meter Test Mode and a prover station are used to confirm the accuracy of the meter and FDM is used to recalibrate the meter if necessary. The Intelis Gas Meter cannot be programmed (configured) or calibrated while the meter is in Test Mode. You must exit Test Mode to program or configure the meter.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter To Adjust the Calibration of an Intelis Gas Meter Before you perform the Adjust Calibration command, you are required to run the meter on a prover station. This determines the meter's accuracy, and the results are used during the Adjust Calibration workflow. Note: The meter must be out of test mode to perform the Adjust Calibration command.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter A sample prover station test result is displayed below to demonstrate an example of adjusting the calibration. Note: Depending on the meter's firmware version, FDM may accept flow rate entries only in a whole number format or it may accept flow rate entries of up to two decimal places. 7. Use the following settings to adjust the meter calibration as it relates to your test results: Prover station mode.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter Valid values: – Cubic Meters Per Hour. 0.01-11.32 – Cubic Feet Per Hour. 0.01 to 400.00 – Station Mode of % Error. -7.00% to +7.00% – Station Mode of % Proof. 93.00% to 107.00% – Station Mode of % Accuracy. 93.00% to 107.00% Note: The Prover flow rate check value must be lesser than the Prover flow rate open value. Prover percentage check. Enter the value from the test mode screen for the meter's check result.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter – % Proof. 100% open, 100% check – % Accuracy. 100% open, 100% check – % Error. 0% open, 0% check 10. Click Quit or Finish to return to the Tools screen. 11. (Recommended) Rerun the meter on the prover station to verify the meter calibration. It is expected that you see the results that match the following example: If the accuracy results are not as expected, repeat the calibration steps.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter The Seal Meter Operation Successful screen indicates that the meter is sealed. 08 July 2020 815-0154-00 REV 000 Itron, Inc.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter 4. Select Finish to return to the Tools screen. To Unseal an Intelis Ggas Meter's UMU 1. From the Tools menu, select OWR Intelis Gas Riva. 2. Enter the Endpoint ID. 3. Select the Endpoint Mode. 4. Select Next. 5. Select Test Mode Operations. 6. Select the Utility ID. 7. Click Next. 8. Select Adjust Calibration for a device with a Seal State of Sealed.
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter 10. (Moblie endpoint mode only) Enter the Unlock ID. 11. Select Next. The Adjust Calibration screen appears. 12. Click Quit or Finish to return to the Tools screen. 08 July 2020 815-0154-00 REV 000 Itron, Inc.