4. Select Finish to return to the Tools screen.
To Unseal an Intelis gas Meter's UMU
1. From the Tools menu, select OWR Intelis Gas Riva.
2. Enter the Endpoint ID.
3. Select the Endpoint Mode.
4. Select Next.
5. Select Test Mode Operations.
6. Select the Utility ID.
7. Click Next.
8. Select Adjust Calibration for a device with a Seal State of Sealed.
A warning message appears indicating that Adjusting Calibration will unseal the meter.
9. Select Yes.
11 December 2019 TDC-1782-005 76
Intelis Gas Meter Installation Guide 6 Adjusting the Calibration