6. After the Intelis Gas Meter is clamped to the prover, turn on the appropriate nozzles in the
center of the screen to exercise the meter.
Note: Typically, Nozzle Number 2-245 CFH is used to exercise the Intelis Meter.
Exercise the meter for 10 to 15 minutes prior to proving the meter.
7. Select Close Nozzles.
8. Unclamp the meter.
9. Exit Special Functions.
Setting Up the Prover
1. If you have not previously done so, log on to the prover.
Note: If you are proving an Intelis Gas Meter for the first time on the prover or if you are
changing proving parameters for the Intelis Gas Meter, you must log on as a Supervisor.
To set up Itron Intelis Gas Meter and ITRON-RTY for the first time, follow steps 2 through
10.
11 December 2019 TDC-1782-005 32
Intelis Gas Meter Installation Guide 5 Proving