4. Select Finish
to return to the Tools screen.
To unseal an Intelis gas meter's UMU
1. From the Tools menu, select OWR Intelis Gas Riva.
2. Enter the Endpoint ID.
3. Select the Endpoint Mode.
• Mobile
4. Select Next.
5. Select Test Mode Operations.
6. Select the Utility ID.
7. Click Next.
8. Select Adjust Calibration for a device with a Seal State of Sealed.
A warning message appears indicating that Adjusting Calibration will unseal the meter.
9. Select Yes.
10. (Moblie endpoint mode only) Enter the Unlock ID.
Adjusting the Intelis Gas Meter Calibration
Intelis Gas Meter Installation Guide TDC-1782-002 60