Note: Logging on to the EEI prover application requires maintenance level security
.
2. Select Maintain > Sensor Maintenance.
3. Clamp the Intelis Gas Meter to the prover and turn on appropriate nozzles in the center
column of the Sensor Maintenance window to exercise the meter
.
Note: Nozzle 192 CFH is typically used to exercise the Intelis Gas Meter.
4. Exercise the meter for 10 to 60 minutes until the Out T
emp value does not vary by more
than 0.2° F within a minute.
5. Close the nozzle.
Proving the Intelis Gas Meter