User's Manual
Electronic Instrument Installation
2. Set the following Default Setup Group parameters:
Enter the desired Cor. Volume Unit parameter. Set Cor. Pulse Volume to the same number of
units.
Enter the desired Unc. Volume Unit parameter. Set Unc. Pulse Volume to the same number of
units.
For Output Pulse Spacing, enter the largest number over 285mS that works for your application.
(Itron recommends a setting of 350mS.)
For Output Pulse Width, enter the largest number over 28mS that works for your application.
(Itron recommends a setting of 40mS.)
3. Click Send at the bottom of the page.
4. Select Current Page.
Wiring
Romet offers different cables and setup options. These instructions include the two most common setup
configurations. For specialized setup instructions, contact Romet.
Military connection. Connect the Itron 100G series remote module using the standard Romet cable (part
number 43-035-40).
Standard Romet 43-035-40 Cable Wiring Table
Corrected count
Uncorrected count
Connection
Romet cable
ERT wire
Romet cable
ERT wire
Pulse Output 1+
Green
White Red White
Blue Blue
Pulse Output 1- White Red Black Red
Note This wiring configuration will not allow a cut cable tamper.
TDC-0824-014 100G Series Gas ERT Module Installation Guide, Remote Mount 72