User's Manual

Electronic Instrument Installation
20. Enter the following parameters in the Corrected Pulse Output Setup section in the lower left of the
window.
Primary Pulse Output. Select K1.
Optional Pulse Output. Set to "Disable".
Pulse Out On ms 70mS
Pulse Out Off ms 500mS AMR
Current/Test/Pulses to Output. Select the number of pulse output pulses to send as a connection test
for the remote ERT module.
Out Cubic Unit/Pulse. Enter the same setting as the ERT module. For example, if the ERT
module consumption is in MCU (thousand cubic feet), set Out Cubic
Unit/Pulse to 1000.
21. Click Send All Changes.
22. Verify all parameters are completed and correct.
23. Click Disconnect.
100G Series Remote Module Installation to the Romet AdEM Using a
Cannon or PG9 Factory-Installed Pigtail Cable
Romet™ AdEM
®
-S/-T/-PTZ 100G Series Remote using the Romet™ Meter Mounting Bracket
This section describes remote 100G ERT module installation with the Romet AdEM
®
-S, AdEM
®
-T and
AdEM
®
-PTZ series correctors.
TDC-0824-014 100G Series Gas ERT Module Installation Guide, Remote Mount 69