17. Slide the encoder down along the side of the encoder spacing tool until it touches the surface of the index
(as shown below). Using moderate pressure, hold the encoder firmly against the index, without moving,
for 15 seconds to permanently apply the encoder.
The photo below shows how the encoder spacing tool and encoder will look after the 15 second wait time.
18. Remove the encoder spacing tool and lay the index on a flat, horizontal surface, to diminish strain on the
encoder cable.
19. Program the index reading (with the encoder mounted) into the 100G Datalogging FN remote ERT
module.
20. Read the 100G Datalogging FN remote ERT module If this reading is the same as the reading
programmed into the 100G Datalogging FN remote ERT module, the ERT module is programmed
correctly.
Index Cover Installation Required Materials
Use the correct replacement index cover gasket for your index (see Replacement Gaskets on page 53 for Itron
4-hole, 2-hole, or 1-hole gasket part numbers).