Owner manual

7.3
3. Highlight the button next to the analyte
groupings to be displayed for selection with
each CC test. In the example shown, the user
would have the choice of selecting ABG (pH,
pCO
2
, pO
2
) and/or Hct/Lytes (Na
+
, K
+
,
iCa, Hct). Press
next to accept entry and go to setup for the next
established product type (if applicable).
Define which analytes will be reported for other cartridge types (if
applicable):
1. The procedure is the same as for CC setup,
with the exception that there are no analyte
grouping choices for User Selects setup; the
user will have the choice of selecting the
individual analytes available on that product
type with each test. Press
next to complete
entry.
LOT ENTRY
QA Users can enter a new product lot and Cal Code into the analyzer memory
using the Lot Entry option. Cal Codes from previously entered product lots
can also be edited via this option.
Entering a new product lot
1. Highlight Lot Entry and press edit.
2. Highlight the appropriate product type and
press
next.
3. Highlight
new in the picklist. Press edit. Up
to 3 lots for each product type can be entered;
when the maximum of 3 has been reached, the
oldest lot for that product type will be deleted.
4. Enter the product lot code and press
next.