Owner manual

2.14
Note:
If Patient ID entry is “Required”, and a patient ID was not entered during
the Calibrating phase, the Patient ID screen will display instead of the
Analyzing screen after sample injection. A patient ID must be entered
before results will display or print. Enter a patient ID and press next to
advance to the Analyzing (or Results) screen.
If the analysis completes before the patient ID is entered, a flashing
“ANALYSIS COMPLETE” message displays in the upper right corner
of the Patient ID screen. Results will not display or print until a Patient ID
is entered and next is pressed. This screen cannot be bypassed; the only
way to exit this screen is to enter a Patient ID.
If a Patient ID has not been entered (no results were displayed or printed),
and IRMA TruPoint powers-down, the test results may be viewed and
printed by recalling the result. If the test being recalled was the last
patient test performed on the analyzer, a patient ID can be entered upon
recall. Refer to Section 4-Recall Patient Results for instructions on
recalling last test result.
If the test being recalled was not the last patient test performed on the
analyzer, the results can be viewed or printed, but a patient ID cannot be
entered. Refer to Section 4-Recall Patient Results for instructions on
recalling test results.
A configuration option may have been selected by the QA User which
defines the minimum and maximum number of characters that must be
entered for a valid Patient ID. If the ID entered does not conform to the
defined ID length, an Invalid Length message will display. Enter a Patient
ID of acceptable length.
MDRD GFR
If a CR test is being performed and the QA User has selected the GFR
calculation settings, the user can enter information to obtain the GFR or
they can skip the GFR.
Enter the following information when the Gender screen appears:
> Select Male, Female or skip the GFR.
Press
next.
> If a gender is selected then enter age in years.
Press
next.