Owner manual

2.2
CAPILLARY REQUIREMENTS
Capillary samples must be collected in the IRMA TruPoint Capillary
Collection Device. See Appendix F for detailed Capillary Collection Device
instructions.
SAMPLE SIZE
The minimum whole blood sample volumes that must be injected into a
single-use cartridge for a patient sample analysis are:
200 uL if sample collected in a syringe
125 uL if sample collected in an IRMA TruPoint Capillary Collection
Device.
Ensure that sufficient sample is collected to meet the minimum volumes
required for injection into the cartridge.
GENERAL SAMPLE COLLECTION GUIDELINES
Time sample collection to minimize delays between collection and
analysis.
Avoid drawing samples above an IV line to prevent dilution of the sample
with IV fluid.
When drawing a sample from an indwelling line, back-flush and clear the
line of IV fluids prior to sampling to remove anticoagulants or medications
which might interfere with the test.
Allow the blood collection site to dry after being cleansed with alcohol to
prevent hemolysis.
Fill the collection device to the appropriate capacity. Incomplete filling
may cause high heparin to blood ratios which may lower ionized calcium
results and may affect other results.
Thoroughly mix samples collected in syringes.
BLOOD GAS SAMPLE HANDLING
Expel any air present in the syringe immediately after collection and
before the sample is mixed. If a portion of the sample must be separated
for other testing, do not expose sample to air.
If a sample cannot be tested within 5 minutes of collection:
> Expel all air from the syringe.
> Cap or seal-off the end of the collection device.
> Store the blood gas sample in an ice slurry.
> Thoroughly mix the sample while the cartridge is calibrating.