Owner manual

D.3
ACCURACY - CONTINUED
Glucose (Glu)
Number of samples: 37
Range evaluated (mg/dL): 19 -338
Slope: 0.970
Intercept: 5.47
Correlation Coefficient (r): 0.992
Sy.x: 11.08
Creatinine (Cr)
Number of samples: 110
Range evaluated (mg/dL): 0.5 - 10.1
Slope: 0.940
Intercept: 0.16
Correlation Coefficient (r): 0.981
Sy.x: 0.55
Lactate (LA)
Number of samples: 117
Range evaluated (mmol/L): 1.1-11.5
Slope: 0.959
Intercept: -0.032
Correlation Coefficient (r): 0.98
Sy.x: 0.46