Owner manual

D.2
ACCURACY - CONTINUED
Potassium (K
+
)
Number of samples: 137
Range evaluated (mM): 1.5 - 13.8
Slope: 0.961
Intercept: -0.033
Correlation Coefficient (r): 0.999
Sy.x: 0.217
Ionized Calcium (iCa)
Number of samples: 137
Range evaluated (mM): 0.80 - 5.70
Slope: 1.019
Intercept: -0.063
Correlation Coefficient (r): 0.997
Sy.x: 0.136
Chloride (Cl
-
)
Number of samples: 56
Range evaluated (mM): 74.8 - 136.4
Slope: 0.956
Intercept: 4.64
Correlation Coefficient (r): 0.986
Sy.x: 3.0
Hematocrit (Hct)
Number of samples: 137
Range evaluated (%): 22.0 - 60.5
Slope: 0.883
Intercept: 2.645
Correlation Coefficient (r): 0.978
Sy.x: 2.45
Blood Urea Nitrogen (BUN)
Number of samples: 56
Range evaluated (mg/dL): 12.5 - 74.6
Slope: 0.985
Intercept: 0.46
Correlation Coefficient (r): 0.995
Sy.x: 1.9