Owner manual

7.25
b3. The Subnet Mask screen displays. This
screen defines which bits are
significant. The default setting is
“255.0.0.0”; other common settings are
“255.255.0.0” and “255.255.255.0”.
Enter the appropriate setting and press
next.
b4. The Remote IP Address screen
displays. Enter the Remote IP Address
using the format “aaa.bbb.ccc.ddd”,
where “aaa”, “bbb”, “ccc”, and “ddd”
are each a 1 to 3 digit value ranging
from 0 to 255. Press
next.
b5. The Remote Port screen displays. This
screen sets the IP Port of the remote
idms computer. The default setting is
“3001”. Enter the appropriate setting
and press
next. Go to step 5.
5. There are two IRMA TruPoint test Results Transfer settings:
Off: The IRMA TruPoint Results Transfer feature is disabled.
On: IRMA TruPoint results may be transferred to idms or other connected
system via serial port, LAN 10/100 or modem. All results transfers utilize
the “ASTM standard 1394-91: Standard Specification for Transferring
Messages Between Clinical Laboratory Instruments and Computer
Systems”, and “ASTM standard 1381-91: Specification for
Low-Level Protocol to Transfer Messages
Between Clinical Laboratory Instruments and
Computer Systems”.
Highlight the “on” or “off” button and press
next. If “on” selected, go to step 6.
6. Highlight “idms” if IRMA TruPoint results
will be transferred to idms, or “host” if results
will be transferred directly to a host system
(e.g., LIS, HIS). Press
next to setup Automatic
Results Device Transfer.