User Manual

7
Menu Action Key Commands
PRINT MENU PRINT
(First) PRINT MENU
(Successive) 1–Print Last
2–Print Pat DB
3–Print QC DB
4–Print by PID
5–Print Check
6–Enter Header
7–Print System
MAIN MENU 0 (Zero)
(First) MAIN MENU
(Successive) 1–Power Level
2–System Time
3–System Date
4–Shutdown
5–System Info
Instrument Shutdown
To shutdown the HEMOCHRON Jr. Signature+, press 0 (zero) to display the main menu and then press 4.
Or, press and hold down the START key for four seconds. If not in use, the instrument will automatically
shutdown after five minutes.
Test Cuvettes
Tests are performed with single-use disposable HEMOCHRON Jr. test cuvettes. Each test cuvette contains a
label, a sample well, a test channel containing reagents, an enclosed waste reservoir, and optical
detection windows.
The cuvette label is human- and instrument-readable. The current test is automatically read by the
instrument and displayed for confirmation when the operator inserts the cuvette into the test chamber.
After the cuvette has warmed to 37 ±1.0 °C, ADD SAMPLE and PRESS START are displayed by the
instrument, alerting the operator to place a drop of blood in the sample well and begin the test by pressing
the START key.
Note: Refer to the package insert accompanying the HEMOCHRON Jr. test cuvettes for storage
and handling instructions.
sample well
overflow area
test channel
label
detection
window