To Review Results of the QC Database:
1.
Press 6. The first line of the results for the most current test record is displayed.
01 QCN EQC
2.
To review each additional line of the results, press 0 or 8 to scroll up or down through the record.
3.
To scroll through additional test records, press ENTER (for the next record) or CANCEL (for the
previous record).
To Exit Database Operations:
1.
Press and hold CANCEL for two seconds.
Acronyms for Error Messages
The following table lists the acronyms for Error Messages when viewing stored Patient or QC records:
Acronym Description Display Message
ORH Out of Range High Out of Range-Hi
ORL Out of Range Low Out of Range-Lo
SNS Sample Not Seen Sample Not Seen
STS Sample Too Small Sample Too Small
STL Sample Too Large Sample Too Large
TTO Temp Timeout Heater Too Cool
TTH Temp Too High Heater Too Hot
SPF Sample Position Fault Sample Pos Fault
Downloading QC/Patient Data to Your PC
For information on downloading QC/Patient data to your PC without ITC data management software, contact
ITC Technical Service for instructions.