Cycle Selection
Table 1: StatSpin MP Cycle Settings. The operator should experiment with different settings to achieve optimum
performance for specific applications. The following are general guidelines:
Setting RPM/RCF Time Rotor
Urine 9,800/3,900 45 seconds RT12
Hematocrit 16,000/13,700 120 seconds RH12
Tube Rotor 15, 800/12,000 30 seconds RT12/RM02
PlasmaRotor
®
20,000/11,200 30 seconds RD01
LipoClear
®
15,800/12,000 95 seconds RT12
Coag Plasma 20,000/11,200 120 seconds RD01
Instructions for Use
1. Lift cover and install rotor.
2. Close and latch cover.
3. Choose desired cycle by depressing the “SET” button until the appropriate LED is illuminated.
4. Press start button.
5. Upon completion of the cycle, the rotor decelerates to a complete stop in 10 seconds and the latch
interlock automatically unlocks.
6. Squeeze the black latch pieces together to open cover.