StatSpin MP Multipurpose Centrifuge Operator’s Manual
Operator’s Manual StatSpin MP Multipurpose Centrifuge Model Number M901 FOR IN VITRO DIAGNOSTIC USE SSMP SSMP-1 SSMP-2 Product Number StatSpin MP for 100-240 VAC, 50/60 Hz (with 2 rotors, RT12 and RH12) StatSpin MP for 100-240 VAC, 50/60 Hz (with 1 rotor, RT12) StatSpin MP for 100-240 VAC, 50/60 Hz (with sample pack of PlasmaRotors) StatSpin is a registered trademark of StatSpin Inc., a Subsidiary of IRIS International, Inc. Copyright 2004 Printed in U.S.A.
i Table of Contents How to use this manual .................................................................................................................. i Section 1 ............................................................................................................................................ 1 Unpacking and Installation ........................................................................................................... 1 Inspect Packaging ..........................................
How to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the information you need to operate and maintain the StatSpin MP. Notes appear in italics to highlight information. When the information requires special attention, a caution symbol appears next to the italicized text.
Section 1 Unpacking and Installation Inspect Packaging The StatSpin MP and its accessories are delivered in one carton. If the centrifuge or accessories have suffered any damage in transport, please inform your carrier immediately. NOTE: Save shipping carton and components to simplify return should service be required. Verify Contents Product No.
Section 2 System Overview Principle and Intended Use IVD For in vitro diagnostic use for rapid separation of whole blood, preparing urine sediment for microscopic analysis and centrifuging microhematocrit tubes for packed cell volume determination. The StatSpin MP is a small, quiet high speed centrifuge. It employs a unique, proprietary drive and suspension system which results in nearly vibration free operation. Light-weight, low mass rotors achieve both top speed and full braking in a few seconds.
Symbols and Definitions (cont.
Accessories Product No. Description Cycle Required RT12 TU15-10 2 x 1.5 ml Fixed Angle Rotor 1.5 ml Pre-calibrated Urine Tube (10 bags of 50) Urine, Tube, LipoClear Urine TP1H 1.3 Lithium Heparin micro centrifuge tube (Bag of 100) Tube Rotor TP1U 1.
Section 3 Operating Instructions Opening and Closing the Cover The centrifuge’s electrically operated cover interlock mechanism prevents operation until the cover is completely closed and latched, and prevents the cover from being opened while the rotor is turning. When the cover is completely closed and locked an operating cycle can be initiated. The centrifuge is also equipped with a manually operated latch that holds the cover down after spinning is complete.
Description of Rotors RT12 Tube Rotor A 2-place rotor designed to accommodate a variety of StatSpin tubes for blood and urine separation, in addition to a variety of standard 1.5 mL and 2.0 mL centrifuge tubes with a maximum diameter of 10.9 mm. These tubes should be supported by the collar of the rotor. CAUTION – DO NOT use glass tubes of any kind in the RT12 rotor.
Cycle Selection Table 1: StatSpin MP Cycle Settings. The operator should experiment with different settings to achieve optimum performance for specific applications. The following are general guidelines: Setting Urine Hematocrit Tube Rotor PlasmaRotor® LipoClear® Coag Plasma RPM/RCF 9,800/3,900 16,000/13,700 15, 800/12,000 20,000/11,200 15,800/12,000 20,000/11,200 Time 45 seconds 120 seconds 30 seconds 30 seconds 95 seconds 120 seconds Rotor RT12 RH12 RT12/RM02 RD01 RT12 RD01 Instructions for Use 1. 2.
Section 4 Maintenance Overview StatSpin recommends that instrument operators perform periodic inspections and preventative maintenance on all StatSpin instruments. Contact StatSpin’s customer service department or distributor if, at any time, the instrument is not functioning properly. CAUTION - Unplug the StatSpin MP from the wall outlet before performing maintenance.
Service Refer all service to qualified service personnel. Reference the StatSpin Warranty for further instruction. Be sure to complete and return the warranty card as directed. Decontamination before returning for service Any instrument or accessory containing accumulated blood and/or other biological or chemical deposits must be cleaned prior to shipment to the manufacturer/dealer for service.
Section 5 Specimen Processing Venous Blood for Coagulation and Chemistry ® PlasmaRotor (RD01) consumable provides a means for the rapid separation of plasma from anticoagulation whole blood, up to 3 mL. To prepare plasma for chemistry testing use the PlasmaRotor cycle and for Coagulation testing using the Coag Plasma cycle. Preparation of Urine Sediment for Microscopic Examination The StatSpin MP quickly prepares urine sediment for microscopic examination.
Procedure 1a. 1b. 2. 3. 4. 5. 6. 7. 8. 9. 10. Capillary ("fingerstick") blood - prepare a skin site and lance. Use heparinized tubes, Product Number HT9H or HP8H. or Venous blood - take well-mixed anticoagulated blood from a syringe or a vacuum blood collection tube. Use untreated tubes, Product Number HT9U or HP8U. Hold the micro-capillary tube by the end with the color-coded band. (See Figures, page 12) Fill to the color-coded band.
Illustrations 12
Appendix Appendix A - Specifications Product No. Model No.
References 1. NCCLS. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Second Edition.” NCCLS document M29-A2 [ISBN 1-56238-453-8]. NCCLS, 940 West Valley Rd, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2001. 2. CDC. Recommendations for prevention of HIV transmission in health care settings. MMWR (Suppl. No. 2S):2S-18S, 1987. 3. CDC.
StatSpin Warranty StatSpin, Inc., a Subsidiary of IRIS International, Inc. warrants that the instruments shall be free from defects in material and/or workmanship, under normal use and service, for the period expiring twelve (12) months from the date of installation, provided the purchaser has completed and forwarded to StatSpin the Warranty Registration Card. StatSpin will, at its discretion repair or replace any unit covered under this warranty returned to StatSpin with shipping costs prepaid.