StatSpin® Express 4 Horizontal Centrifuge Operator’s Manual A Division of IRIS International, Inc.
Operator’s Manual StatSpin® Express 4 Horizontal Centrifuge Model Number M510 For in-vitro diagnostic use Product Number ® SSH4 – StatSpin Express 4 for 100-240 VAC, 50/60 Hz StatSpin and Iris are registered trademarks of IRIS International, Inc. Copyright 2010 Printed in U.S.A.
Table of Contents How to use this manual…………………………………………..……….………… ii Section 1 ……………………………………………………………..……….……….. 1 Unpacking and Installation ..………………………………………………………... 1 Inspect Packaging ………………………………………………………………...………. ................. 1 Verify Contents ………………………………………………………………...………......................... 1 Install ………………………………………………………………...………. ........................................... 1 Connect Power ………………………………………………………………...………......................... 1 Section 2 ……………………………………………………………..
How to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the information you need to operate and maintain the StatSpin® Express 4. When the information requires special attention, a caution symbol italicized text. appears next to the Please pay close attention to the instructions that accompany the notes and symbols as well as the standard laboratory practices outlined by your facility and local regulatory agencies.
Section 1 Unpacking and Installation Inspect Packaging The StatSpin® Express 4 and its accessories are delivered in a single carton. If the centrifuge or accessories have been damaged in transport, please inform your carrier immediately. NOTE: Save the shipping carton and packing materials to facilitate return should service be required. Verify Contents The package contains: One StatSpin® Express 4 horizontal centrifuge (Product No. SSH4) One centrifuge rotor (Product No.
Section 2 System Overview Principle and Intended Use IVD For in-vitro diagnostic use to produce the rapid separation of whole blood contained in original collection tubes. The StatSpin® Express 4 is a high-speed horizontal bench top centrifuge used to rapidly separate blood components in the original sample collection tubes. Do Not centrifuge substances with a density greater than 1.2 kg/dm3.
Locked Locked position for shield Unlocked Unlocked position for shield Operator Controls Start Stop/Open The Start button initiates the selected cycle. The Stop/Open button interrupts the cycle and releases the cover when the rotor has stopped. 3 min 3 Minute Cycle The 3 min button selects a cycle of three minutes at 5100 rpm. 5 min 5 Minute Cycle The 5 min button selects a cycle of five minutes at 5100 rpm.
Error Codes Table Failure to reach speed in 30 seconds. Confirm no obstruction in bowl area, contact customer service. 3 min 3 min 5 min Latch is not locked. Confirm cover is fully seated in closed position. 5 min Rotor exceeded set speed. Remove and inspect rotor, reinstall. 3 min 3 min 10 min Cover is not closed. Confirm cover is fully seated in closed position. 10 min Out of balance. Confirm tubes are balanced in rotor and centrifuge is on a level surface.
5
Section 3 Operating Instructions Opening and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover is not completely closed and locked, and prevents the cover from opening when the rotor is turning. An operating cycle can be initiated only if the cover latch is engaged. Cover Interlock By-pass In case of power failure or malfunction, the Stop/Open button may not release the cover.
2. Locate 1/8 inch hex key supplied and remove center screw of rotor hub. 3. Remove rotor lifting straight up and out of centrifuge. To re-install the rotor: 1. 2. 3. 4. Align position 1 on the rotor with the flat on the motor shaft. Slide the rotor onto the shaft and push down gently to seat. Do not force the rotor onto the shaft. If it does not slide down easily re-align the rotor with the shaft. Reinstall center screw and tighten with supplied hex key. Reinstall shield into housing.
Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenance on all instruments. Contact customer service or your distributor if the instrument is not functioning properly. CAUTION - Unplug the StatSpin® Express 4 from the wall outlet before cleaning or performing maintenance. WARNING – Inspect instrument for cracks or any physical damage to housing, cover and rotor.
Periodic Maintenance Check When the cycle counter in the instrument reaches 50,000 spins, the Error/Service indicator will continuously illuminate. This is a reminder to perform some important maintenance checks on the instrument. Follow these steps to complete the maintenance check and reset the cycle counter: 1. Remove protective shield and rotor from the instrument (see Section 3 – Removing and Installing the Rotor). 2.
Troubleshooting Symptom: Solution: Red Error / Service indicator flashing, unit will not spin. Refer to Error Code Table of manual to identify and correct problem. Tubes are loose and do not fit into buckets. Be sure the correct tube inserts are in use. Check insert listing available under accessories to match with appropriate tube sizes. No LEDs are on and unit will not spin. Main power switch is located on rear panel along with fuse holder.
Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. Instrument fuse(s) are located on the rear panel. Remove the line cord to access the fuse holder. Replace with fuses of the same type and rating. 2.5A 250V F 5 x 20 mm Be sure to complete the Warranty Registration Card as directed.
Section 5 Appendix Specifications Product Number SSH4 Model Number M510 Spin Parameters 3 min: 180 seconds @ 5100 rpm / 4000 xg 5 min: 300 seconds @ 5100 rpm / 4000 xg 10 min: 600 seconds @ 4500 rpm / 3100 xg Acceleration Time ≤ 30 seconds Deceleration Time ≤ 30 seconds Electrical 100 – 240 VAC, 50/60Hz, 2.5 Amps Dimensions Height: 8.0“ (20.3 cm) Width: 13.0” (33.0 cm) Depth: 16.0” (40.6 cm) Weight: 20 lbs.
References 1. 2. 3. CLSI. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Third Edition.” CLSI document M29-A3 [ISBN 1-56238-567-4]. CLSI, 940 West Valley Rd, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2005. CDC. Recommendations for Prevention of HIV Transmission in Health Care Settings. MMWR (Suppl. No. 2S):2S-18S, 1987. CDC.
Iris Sample Processing Warranty Iris Sample Processing, a Division of IRIS International, Inc. warrants that the centrifuge, with the exception of the rotor, shall be free from defects in material and/or workmanship, under normal use and service, for the period expiring twenty-four (24) months from the date of installation. The rotor shall be free from defects in material and/or workmanship, under normal use and service, for the lifetime of the instrument.