Owner manual

9
Troubleshooting
No power light
Is the line cord completely plugged
into the wall?
No
Plug line cord into the wall.
Yes
Is the wall outlet working properly?
No
Try another wall outlet.
Yes
Is the fuse blown?
Yes
Replace fuse.
No
Contact Iris Sample Processing or
an Authorized Repair Facility.
Centrifuge will not spin, or shuts
off prematurely
Is the cover closed and latched?
No
Completely close the cover.
Yes
Is the error light flashing?
Yes
Identify source of error from flashing
indicator lights, correct the error condition,
and push the Stop/Open button to clear the
error.
No
Contact Iris Sample Processing or
an Authorized Repair Facility.
Centrifuge will not open at end
of cycle
Is the bowl gasket completely
seated and sealing the bowl?
NO
Press the bowl gasket all the way down into
the housing slot.
YES
Contact Iris Sample Processing or
an Authorized Repair Facility.