E Ex xp prre es ss s 3 Primary Tube Centrifuge Operator’s Manual A Division of IRIS International, Inc.
Operator’s Manual Express 3 Primary Tube Centrifuge Model Number M502-22 For in-vitro diagnostic use Product Number SSX3 - Express 3 for 100-240 VAC, 50/60 Hz StatSpin and Iris are registered trademarks of IRIS International, Inc. Copyright 2006 Printed in U.S.A.
Table of Contents How to use this manual…………………………………………..……….………… ii Section 1 ……………………………………………………………..……….……….. 1 Unpacking and Installation ..………………………………………………………...1 Inspect Packaging ………………………………………………………………...………................... 1 Verify Contents ………………………………………………………………...………. ........................ 1 Install ………………………………………………………………...………. ............................................ 1 Connect Power ………………………………………………………………...………. ........................ 1 Section 2 ……………………………………………………………..
How to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the information you need to operate and maintain the Express 3. Notes appear in italics to highlight information. When the information requires special attention, a caution symbol italicized text.
Section 1 Unpacking and Installation Inspect Packaging The Express 3 and its accessories are delivered in a single carton. If the centrifuge or accessories have been damaged in transport, please inform your carrier immediately. NOTE: Save the shipping carton and packing materials to facilitate return should service be required. Verify Contents The package contains: One Express 3 centrifuge (Model No. M502-22) One centrifuge rotor (Product No. RTX8) Eight rotor inserts for 13x75 mm tubes (Product No.
Section 2 System Overview Principle and Intended Use IVD For in-vitro diagnostic use to produce the rapid separation of whole blood contained in original collection tubes. The Express 3 is a high-speed bench top centrifuge used to rapidly separate blood components in the original sample collection tubes. The eight-place 30° angle rotor included with the centrifuge produces rapid separation while minimizing potential for re-suspension of cells.
Operator Controls Start The Start button initiates the selected cycle. Stop/Open The Stop/Open button interrupts the cycle and releases the cover when the rotor has stopped. The 2 min button selects a cycle of two minutes at 7200 rpm. 2 min 2 Minute Cycle 3 min 3 Minute Cycle 5 min 5 Minute Cycle The 5 min button selects a cycle of five minutes at 5600 rpm.
Accessories Product No. Description SV02 Inserts for 3 mL tubes (10.25 x 64 mm) 4/pk. SV03 Inserts for 2 mL tubes (10.25 x 47 mm or BD Microtainers) 4/pk. SV04 Inserts for 1.5-2 mL Microtubes (e.g. LipoClear) 4/pk. SV05 Inserts for 2 mL tubes (BD Microtainers) 4/pk. SV06 Inserts for 7 mL tubes (16 x 75 mm) 8/pk. SV07 Inserts for 10 mL tubes (16 x 100 mm) 8/pk. Included with centrifuge SV08 Inserts for 5 mL tubes (13 x 75 mm) 8/pk.
Section 3 Operating Instructions Opening and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover is not completely closed and locked, and prevents the cover from opening when the rotor is turning. An operating cycle can be initiated only if the cover latch is engaged. Cover Interlock By-pass In case of power failure or malfunction, the Stop/Open button may not release the cover.
2. Slide the rotor onto the shaft and push down gently to seat the drive pins in the holes. 3. Place the retaining nut on the shaft and tighten securely using the 5/16 inch hex key. You will not be able to tighten the retaining nut if the drive pins are not properly aligned. If this occurs lift the rotor off and repeat the process until it is seated. Spinning the Sample CAUTION - Never operate the Express 3 without the rotor locked securely in place.
Unloading 1. 2. 3. Upon completion of the cycle, the rotor decelerates to a complete stop in 30 seconds or less, a signal beep will sound, and the cover latch automatically releases. Lift the cover. Carefully remove the tubes using care not to re-suspend the sample.
Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenance on all instruments. Contact customer service department or your distributor if the instrument is not functioning properly. CAUTION - Unplug the Express 3 from the wall outlet before cleaning or performing maintenance.
Troubleshooting No power light ↓ Is the line cord completely plugged into the wall? ↓ Yes ↓ Is the wall outlet working properly? ↓ Yes ↓ Is the fuse blown? ↓ No ↓ Contact Iris Sample Processing or an Authorized Repair Facility. → No→ Plug line cord into the wall. → No→ Try another wall outlet. → Yes→ Replace fuse. → No → Completely close the cover. → Yes → Identify source of error from flashing indicator lights, correct the error condition, and push the Stop/Open button to clear the error.
Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. 1-800-782-8774 Instrument fuse(s) are located on the rear panel. Remove the line cord to access the fuse holder. Replace with fuses of the same type and rating. 3.0A 250V F 5 x 20 mm Be sure to complete the Warranty Registration Card as directed or go to http://www.proiris.com/ to obtain a copy of the warranty card.
Section 5 Appendix Specifications Product Number SSX3 Model Number M502-22 Spin Parameters 2 min: 120 seconds @ 7200 rpm / 4440 xg 3 min: 180 seconds @ 7200 rpm / 4440 xg 5 min: 300 seconds @ 5600 rpm / 2685 xg Acceleration Time ≤ 25 seconds Deceleration Time ≤ 30 seconds Electrical 100 – 240 VAC, 50/60Hz, 3.0 Amps Dimensions Height: 7” Width: 11” Depth: 10” Weight: 8¾ lbs.
References 1. 2. 3. CLSI. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Second Edition.” CLSI document M29-A2 [ISBN 1-56238-453-8]. CLSI, 940 West Valley Rd, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2001. CDC. Recommendations for Prevention of HIV Transmission in Health Care Settings. MMWR (Suppl. No. 2S):2S-18S, 1987. CDC.
Iris Sample Processing Warranty Iris Sample Processing, a Division of IRIS International, Inc. warrants that the instruments shall be free from defects in material and/or workmanship, under normal use and service, for the period expiring twenty-four (24) months from the date of installation, provided the purchaser has completed and forwarded to Iris Sample Processing the Warranty Registration Card.