User Manual
3
Symbols and Definitions (cont.)
REF
Product/Reference
Number
Indicates the StatSpin product/catalog number
Caution Statement of caution/warning, read instruction
carefully
Temperature
limitation
Indicates storage requirements range
EC Representative European Community Authorized Representative
IVD
For in vitro
diagnostic use
Clarifies for use as in vitro diagnostic only
Non sterile Indicates non-sterile product
S N
Serial Number Indicates instrument serial number code
Consult
Instructions
Consult instruction manual or insert sheet for further
explanation
Error Indicators
Error codes for the Express 2 are specified by a combination of the Error/Service LED flashing or
continuous and any combination of the cycle LEDs on the front panel.
Error/Service = Quick Spin = 30 s Normal Spin = 120 s Hard Spin = 180 s
Error/Service light will continuously illuminate when the centrifuge has achieved a total cycle count of 18,000, which
is the useful life of the drive system. Drive mechanism needs replacing. Contact an authorized service center.
Accessories
Product No. Description
SV01 Inserts for 5 mL tubes (13 x 75 mm or 10.25 x 82 mm) 4/pk. Included with centrifuge
SV02 Inserts for 3 mL tubes. (10.25 x 64 mm) 4/pk.
SV03 Inserts for 2 mL tubes. (10.25 x 47 mm or BD Microtainers) 4/pk.
SV04 Inserts for 1.5-2 mL Microtubes (e.g. LipoClear) 4/pk.
DL01 Disposable chamber liners. 10/pk.
RTX4 Express 2 Rotor. Includes four SV01 inserts.
Error/Service
light flashing
or continous
Rotor failed
to reach rpm
within 30 s
Cover opened
or cycle
changed during
operation
Insufficient
power to
maintain
rpm
Rotor spun
backward
direction during
reverse braking