Instruction Manual

8
Troubleshooting
No Power Light
Is the power cord completely
plugged into the wall?
NO
Plug power cord into the wall.
YES
Is the wall outlet working properly?
NO
Try another wall outlet.
Is the barrel connector from the
transformer completely plugged into
the centrifuge?
NO
Plug the barrel connector into the centrifuge.
YES
Contact StatSpin or an Authorized
Repair Facility.
Centrifuge Will Not Spin/Shuts
Off Prematurely
Is the cover completely closed and
cannot be opened manually?
NO
Completely close the cover.
YES
Contact StatSpin or an Authorized
Repair Facility.
Centrifuge will not open at end
of cycle
Is the bowl gasket completely
seated and sealing the bowl?
NO
Press the bowl gasket all the way down into
the housing slot.
YES
Contact StatSpin or an Authorized
Repair Facility.