Owner's manual
3
Symbols and Definitions (cont.)
REF
Product/Reference
Number
Indicates the StatSpin product/catalog number
Caution Statement of caution/warning, read instruction
carefully
Temperature
limitation
Indicates storage requirements range
Do not re-use
Indicates product is single use only
EC Representative
European Community Authorized Representative
IVD
For in vitro
diagnostic use
Clarifies for use as in vitro diagnostic only
Non sterile Indicates non-sterile product
LOT
Batch code Indicates product lot number
S N
Serial Number Indicates instrument serial number code
Use by
Indicates product expiration date
Consult
Instructions
Consult instruction manual or insert sheet for further
explanation
Audible Indicators
Normal Function Codes
Sound Meaning
2 medium beeps Sounds on power up; instrument ready
3 short beeps Cycle completed per specified operating parameters.
Chirp
Start or stop command recognized by the
microprocessor.
Malfunction Codes
1 long beep followed by: 1 short beep Rotor failed to reach selected rpm within allotted time
1 long beep followed by: 2 short beeps Lid opened during operation
1 long beep followed by: 3 short beeps
Internal temperature exceeded 60°C (unit can not be
restarted until internal temperature drops below 52°C)
1 long beep followed by: 4 short beeps Insufficient power to maintain rpm
1 long beep followed by: 5 short beeps Over current detector tripped.
Cell and Filter Concentrators
A variety of concentration options are available for use with the CytoFuge 2. See the package
inserts for the Filter Concentrators (both reusable and disposable) and the Cell Concentrators (both
reusable and disposable) for detailed information.