Owner's manual

2
Section 2
System Overview
Principle and Intended Use
For in vitro diagnostic use in concentrating cells onto a microscope slide.
The CytoFuge 2 is a microprocessor-controlled cell preparation system that uses centrifugal force to
deposit cells onto microscope slides. Samples are centrifuged in reusable or disposable gasket-
sealed chambers (Cell Concentrators), or in disposable chambers (Filter Concentrators) that include
a filter material to absorb and capture suspension fluid during cytocentrifugation
A four-position rotor with transparent cover is supplied as standard equipment. The system is
designed to be operated on a laboratory bench or, if potentially biohazardous material is processed,
in a biological safety cabinet. A cover interlock system prevents operation until the centrifuge cover
is securely latched, and prevents access to spinning parts until the rotor has come to a complete
stop. Speeds from 600 rpm to 4400 rpm and cycle times from 2 minutes to 30 minutes may be
selected. Automatic acceleration rate control and dynamic braking are included to protect delicate
samples.
Symbols and Definitions
Start button The start button initiates a pre-timed cycle at the
speed selected. NOTE: The CytoFuge 2 does not
have an on/off switch and is normally left plugged in
and “on”.
Stop/open button
The stop/open button interrupts the cycle and stops
the centrifugation. This button may also be used to
release the cover.
Power indicator
The green LED identified as power is illuminated
when the CytoFuge 2 is connected to a power source.
Release cover
Located below stop/open button on instrument front
panel
Time selector Available time settings: 2, 4, 6, 8, 10, and 30 minutes
Speed selector
Available speed settings: 600, 700, 850, 1000, 1300,
1600, 2200, and 4400 rpm
Power input DC Power Input: 24V DC Plug-in
IVD