Manual

3
Error Indicators
Error codes for the CritSpin are specified by the Error/Service LED flashing or continuously
illuminated.
Error/Service =
Error/Service light will continuously illuminate when the centrifuge has achieved a total cycle count of
18,000, which is the useful life of the drive system. Drive mechanism needs replacing. Contact an
authorized service center.
Error/Service light will flash when one of the following conditions occur:
1. System Error – Restart the centrifuge.
2. Motor Failure – Drive mechanism requires replacement.
3. Rotor Improperly SeatedRe-install rotor.
Error/Service and 120 s lights will flash when the cover is open during spin.
Accessories
Product No. Description
RH12 Hematocrit rotor, 12-position
HR4C Circular Hematocrit Reader
HR05 Card-style Hematocrit Reader
HT9H-10 Hematocrit Tubes, 40mm, glass, heparin
HT9U-10 Hematocrit Tubes, 40mm, glass, untreated
HP8U-10 Hematocrit Tubes, 40mm, plastic, heparin
HP8U-10 Hematocrit Tubes, 40mm, plastic, untreated
HS24-10 Clay Sealant Pad
CSD1 Digital Reader (included with CS22 Product Number)
A
B
C
120S
A Error/Service light
continuously
illuminated
B Error/Service
light flashing
120S
C Error/Service and
120 s flashing