User's Manual

User Interface
29 Canis014D07 User Manual
Exposure Time Setting
You can enter the independent exposure mode by clicking the [application] button of the setting page and selecting the
independent mode.
The exposure time can be controlled by clicking the ExposureTime increase or decrease on the screen, and the
exposure time can also be adjusted through the physical buttons of the equipment head. exposure is performed by long
pressing the 1S exposure button, and the exposure conditions are battery voltage, power level, tube temperature, and
time interval. Only when certain conditions are met, a buzzer will be heard during exposure, and the end of the buzzer
indicates the end of the exposure. If the exposure button is released during the buzzer sound, the exposure will end at
any time.
Press and hold the exposure button on the equipment for 1 second to expose exposure. When certain conditions are
met for battery voltage, power, tube temperature, time interval, and detector status, exposure can be performed. During
exposure, you will hear a buzzer sound, and the end of the buzzer indicates the end of the exposure. If you release the
exposure button during the buzzer sound, the exposure will end at any time.
You can change the exposure judgment adjustment by modifying the configuration file, such as the range of exposure
temperature, the range of battery power, the range of battery voltage, the interval of exposure time, the voltage and
current during exposure;
The path to the configuration file is in Explorer /iray/default.config. Open the file with an editor, edit and save it. The
configured parameters are kv, ua, minTemp, maxTemp, minBattery, maxBattery, minVoltage, maxVoltage, isMarket.
Among them, when isMartket is set to false, the time interval of two exposures is 30 times, and when it is set to true,
the time interval of two exposures is 7 times. The relevant configuration file page is shown in the figure below