Users Manual

6. Regulatory Information
User Manual of Mars1417X
75
6.2. Guidance and Manufacture’s Declaration for EMC
6.2.1. EMI Compliance Table
Emissions
Phenomenon Compliance Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class
B
Professional healthcare facility
environment
Harmonic
distortion
IEC 61000-3-2
Class A
Professional healthcare facility
environment
Voltage
fluctuations and
flicker
IEC 61000-3-3
Compliance
Professional healthcare facility
environment
6.2.2. EMS Compliance Table
Enclosure Port
Phenomenon Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM
field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Near fields from
RF wireless
communications
equipment
IEC 61000-4-3
Refer to table “Near fields from RF
wireless communications equipment”
Rated power
frequency magnetic
fields
IEC 61000-4-8 30A/m
50Hz or 60Hz
Near fields from RF wireless communications equipment
Test
frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m