Users Manual Part 2

4. Software Setup
User Manual of Mars1417X
55
Upload Lag Upload Lag into SDK current directory
Active Select one item in list. Click this button to activate selected
template for hardware correction.
UpdateHWPreOffset Force detector update Offset template(not needed for
postoffset flow)
ReadStatus Get the current state of template for hardware correction,
enable/disable
Generate Gain Template File
If the relative position between tube and detector changed or KV value changed, it suggest to
create gain template file.
1. Enter Create Gain page
Click "Start" button to start process, the offset type should be selected, then start to get the
images.