Users Manual Part 2

7. Trouble Shooting
User Manual of Mars1417X
71
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
Input a.c. power port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5 ±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
Voltage dips
IEC 61000-4-
11
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and
315º
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0º
Voltage
interruptions
IEC 61000-4-
11
0% UT; 250/300 cycles
Recommended separation distances between portable or mobile RF communication
device and detector:
Portable RF communications equipment, including antennas, can effect medical electrical equipment.
The warning should include a use distance such as “be used no closer than 30 cm (12 inches) to any part of
the [ME EQUIPMENT or ME SYSTEM], including cables specified by the manufacturer”.
Cable provided for EMC
Cable
Recommended
length
Shielded/Unshielded Number
Cable
classification
AC power cable 1.8m Unshielded 1 piece AC power
DC power cable 3m Shielded 1 piece DC power
Ethernet cable 3.5m Shielded 1 piece Signal