Users Manual Part 2

6. Regulatory Information
User Manual of Mars1417X
75
6.2. Guidance and Manufacture’s Declaration for EMC
6.2.1. EMI Compliance Table
Emissions
Phenomenon Compliance Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class
B
Professional healthcare facility
environment
Harmonic
distortion
IEC 61000-3-2
Class A
Professional healthcare facility
environment
Voltage
fluctuations and
flicker
IEC 61000-3-3
Compliance
Professional healthcare facility
environment
6.2.2. EMS Compliance Table
Enclosure Port
Phenomenon Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM
field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Near fields from
RF wireless
communications
equipment
IEC 61000-4-3
Refer to table “Near fields from RF
wireless communications equipment”
Rated power
frequency magnetic
fields
IEC 61000-4-8 30A/m
50Hz or 60Hz
Near fields from RF wireless communications equipment
Test
frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m