Users Manual Part 2
Wireless Digital Wireless Digital Flat Panel Detector Mars1417X
58 User M
anual of Mars1417X
5.3.3.
IDLE
XWIN1
(fixed)
X‐Ray
imagin g
XWIN1
(fixed)
Offset
imagin g
X‐Ray Info
RAWPreview
RAWX‐Rayim ag e
(IfHW‐offs etisdisabled)
RAWOffset(Orcorrected)
image
1. For Freesync mode, there is no low power state.
2. When the detector is Idle, user can start the exposure flow any time.
3. When the X-Ray starts, the detector will sense the X-Ray automaticlly, the XWIN is
configured by parameter “Set Delay Time” on “Detector” page “Parameter” tab, the unit is
“ms”, user needs to make sure that the XWIN is larger than the X-Ray time.
4. After the XWIN is end, then the detector will start the acquisition flow.
5. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will
be sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-
Ray image will not be sent and only the corrected image will be transferred.
5.4. After use
1. Disconnect the software
2. Power off
3. Keep it clean
4. Store under specified conditions
5.5. Firmware Upgrade
On “Detector” Page,“Parameter” Tab,user can upgrade firmware by entrance button