Users Manual Part 2

5. Operation Instructions for Image Acquisition
User Manual of Mars1417X
69
5.3.1. Inner
IDLE
XWIN1
(fixed)
XRay
imaging
XWIN1
(fixed)
Offset
imaging
XRay Info
RAWPreview
RAWXRayimage
IfHWoffsetisdisabled
RAWOff set(Orcorrected)
image
Lowpower
Prep Ack
1. The detector is in low power state, user needs to send Cmd “Prep” to make the detector
exit to idle state which indicated by the acknowledge to Cmd “Prep”.
2. When the detector is in idle state, user can start the X-Ray any time.
3. When the X-Ray starts, the detector will sense the X-Ray automaticlly, the XWIN is
configured by parameter “Set Delay Time” on “Detector” page “Parameter” tab, the unit is
“ms”, user needs to make sure that the XWIN is larger than the X-Ray time.
4. After the XWIN is end, then the detector will start the acquisition flow.
5. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will
be sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-
Ray image will not be sent and only the corrected image will be transferred.
5.3.2. Freesync Mode