6. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will
be sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-
Ray image will not be sent and only the corrected image will be transferred.
5.2.3. Work Flow(Prep+Acq)
IDLE Prepare
XWIN2
(dynamic)
XRay
imaging
Cmd_Prep Ack_Prohibit Ack_Enable
XWIN2
(dynamic)
Offset
imaging
RAWPreview
RAWXRayimage
(IfHWoffsetisdisabled)
RAWoffset(Orcorrected)
image
Cmd_S ingleAcq
XRay
1. Send Cmd“Prep” on UI “Acquire” page.
2. After receiving the Cmd_Prep, it will start the prepare process, and send back the
acknowlage of “Prohibit” and “Enable”, the “XWIN” will be started.
3. The max XWIN is configured by parameter “Exp Window Time” on “Detector” page
“Parameter” tab, the unit is “ms”.
4. User starts the X-Ray.
5. Send “SingleAcq” on UI “Acquire” page after the X-Ray is end.
6. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will
be sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-
Ray image will not be sent and only the corrected image will be transferred.
5.3. AED Mode