5. Operation Instructions for Image Acquisition
User Manual of Mars1417X
67
HVG
X-Tube
FPD
Work
Station
5.2.2. Work Flow(PrepAcq)
Select HWPostOffset、HWGain、HWDefect. If user need the raw image, please de-select all
these correction options.
Also, the software correction is supported.
IDLE Prepare
XWIN1
(fixed)
XRay
imaging
Cmd_PrepAcq Ack_Prohibit Ack_Enable
XWIN1
(fixed)
Offset
imaging
RAWPreview
RAWXRayim ag e
(IfHWoffsetisdisabled)
RAWoffset(Orcorrected)
image
1. Send Cmd“PrepAcq” on UI “Acquire” page.
2. After receiving the Cmd_PrepAcq, it will start the prepare process, and send back the
acknowlage of “Prohibit” and “Enable”, the “XWIN” will be started.
3. The XWIN is configured by parameter “Clear Acq Delay Time” on “SDK” page, the unit is
“ms”.
4. User needs to make sure the X-Ray ends within the XWIN.
5. The detector will send the images after the XWIN closed.