Users Manual Part 2
5. Operation Instructions for Image Acquisition
User Manual of Mars1417X
55
Mars1417X provides SDK for users to integrate detector into their DR system.
Additionally, it also provides an application for demonstration, i.e. IDetector. User can use
IDetector to control detector without DR system.
5.1. Steps for acquiring image
Make sure the hardware is connected correctly and then power on.
Once powered off, please wait at least 60s before power on again
Wait until initialization is complete
Connect the software
choose the synchronization mode
Generate HWPreOffset, Gain and Defect template after the detector reaches thermal
equilibrium
Acquire images in the selected mode
To Acquire X-ray image is the main operation of Mars1417X. Most importantly, detector should
build synchronization with X-ray generator. Mars1417X has one synchronization modes to
acquire X-ray image, which is Software Mode.
5.2. Software Mode
5.2.1. Block Diagram
Software mode is the basic way to acquire X-ray image. Please see figure below for general
feature. Workstation is a host PC device installed with iDetector and SDK. FPD is the Wireless
Digital Wireless Digital Flat Panel Detector and HVG is the High Voltage Generator. In this
mode, Workstation does not have to control X-ray generator. Users would decide when to shoot
X-ray.