1.2. Notes for Using
When using the product, take the following precautions. Otherwise, problems may occur and
the product may not function correctly.
Before exposure
Be sure to check the connection of all the parts are set properly & check the detector is
kept in insulated cover that operator or patient can’t touch the detector directly before
powered up.
Be sure to check the product daily and confirm it work properly.
Sudden heating of the room in cold areas will cause condensation on the product. In this
case, wait until the condensation evaporates before performing an exposure. If it is used
when condensation is formed, problems may occur in the quality of captured images.
When an air-conditioner is used, be sure to raise/lower the temperature gradually to
prevent condensation.
The product should be warmed up for 15 minutes before exposure or updating the gain
map and defect map.
Make sure wave form of the energy going to the X ray tube is square not pulse.
Be cautious with circumstance that someone has radio isotope recently injected into them,
it may cause panel transmit image without x ray.
Once powered off, please wait at least 60s before power on again
During exposure
Do not move Power Cable or Ethernet Cable during exposure, or it may cause image noise
or artifacts, even incorrect images.
Do not use the product near the equipment generating a strong magnetic field. Otherwise,
it may cause image noise, artifacts or even incorrect images.