The correction and calibration should be performed after installation and it is recommended to
perform the new correction and calibration after any major change on the system settings and
hardware configuration. On the other hand, it is also recommended to do the correction and
calibration in each 6 months.
5.5.1. HW pre-offset Template Generation
Enter Acquire interface, select
HWPostOffset option
Enter Calibrate interface, click
UpdateHWPreOffset button.
Waiting until status bar displayed:
"Task succeed:
HwGeneratePreOffsetTemplate”
5.5.2. Gain Calibration Template Generation
If the relative position between tube and detector changed or KV value changed, it suggest to
create gain template file.